47 / 120
E3KOM-7.00
CardioTouch3000 Operation Manual
8) Printing Form
In this section, descriptions and samples of various printing forms are provided.
Rhythm Printing Form
- 3CH printing: To print 3 channels at the same time (I ~ III / aVR ~ aVF / V1 ~ V3 /
V4 ~ V6)
- 6CH printing: To print 6 channels at the same time (I ~ aVF / V1 ~ V6)
- 12CH printing: To print 12 channels at the same time (I ~ V6)
Record Printing Form
- ‗3CH+1‘: I, II and III will be recorded for the first 2.5 seconds, V1~V3 for the next
2.5 seconds, V4~V6 for the next 2.5 seconds and 1 channel of rhythm will be
recorded for 10 seconds at the bottom.
- ‗6CH+1‘: I~ aVF will be recorded for the first 5 seconds and V1~ V6 for the next 5
seconds. And 1 channel of rhythm will be recorded for 10 seconds at the bottom.
- ‗12CH‘: 12-channel rhythm will be recorded as I~ V6 at the same time for 10
seconds.
- ‗3CH+3‘: I~ III will be recorded for the first 2.5 seconds, aVR~ aVF for the next 2.5
seconds, V1~V3 for the next 2.5 seconds and V4~V6 for the next 2.5 seconds. And
3 channels of rhythm will be recorded for 10 seconds at the bottom.
Form of Beat Report Output
- Choose an additional output form after printing out diagnosis; either Text, Guide or
Vector.
- Text: to output representative beat and each diagnostic parameter. Output will come
at 50mm/sec if the output speed is set at ‗5, 12.5 and 25 mm/sec‘, whereas it will come
at 100 m/sec if set at ‗50 and 100mm/sec‘.
- Guide: to output representative beat and each diagnostic parameter, with diagnostic
guide for arrhythmia. Output will come at 50 mm/sec regardless of the output speed.
- Vector: to output representative beat and each diagnostic parameter and Vector of
QRS, as indicated by the arrow. Output will come at 50 mm/sec regardless of the
output speed.
Содержание CardioTouch3000
Страница 1: ...CardioTouch3000 Operation Manual Ver 7 00 2019 05 23 www ebionet com...
Страница 48: ...48 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 49: ...49 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 50: ...50 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 51: ...51 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 12 CH Record Report Form BEAT REPORT TEXT...