Network Setup
When connecting the equipment with LAN to interface with the outside PC, network
setup is needed. Network IP is required and DNS is not used.
When selecting the Network menu at the left side of the System Setup page, Network
Setup page will show up at the right side.
The mode of network connection can be selected as Wired or Wireless in ‗Net Device‘
setup.
The automatic IP(DHCP) or fixed(Manually) can be selected through the 'Configure'.
If the Configure is setup as automatic IP(DHCP), the Device IP, Subnet Mask, Gateway
value is received automatically from the DHCP server on your network. If the Configure is
set to manually, the user has to enter the Device IP, Subnet Mask, Gateway directly.
When selecting one of the ‗Device IP‘, ‗Subnet Mask‘, ‗Gate Way‘, the keypad will show
up. Then, input the information required and press 'OK' button to save or 'ESC' to cancel the
information.
The server interlocking device can be selected as BMS Server of Bionet.
After selecting the type of interlocking server, select the 'Setting' button; an appropriate
setup window to the choice of server will be displayed.
Содержание CardioTouch3000
Страница 1: ...CardioTouch3000 Operation Manual Ver 7 00 2019 05 23 www ebionet com...
Страница 48: ...48 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 49: ...49 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 50: ...50 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 51: ...51 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 12 CH Record Report Form BEAT REPORT TEXT...