Guidance and manufacturer’s declaration – electromagnetic immunity
The CardioTouch3000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the CardioTouch3000 should assure that it is used in such an
environment.
IMMUNITY test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V rms
Portable and mobile RF
communications equipment should be
used no closer to any part of the
CardioTouch3000, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
80 MHz to 800 MHz
800 MHz to 2,5 GHz
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the
recommended separation distance in
meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
Содержание CardioTouch3000
Страница 1: ...CardioTouch3000 Operation Manual Ver 7 00 2019 05 23 www ebionet com...
Страница 48: ...48 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 49: ...49 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 50: ...50 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 51: ...51 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 12 CH Record Report Form BEAT REPORT TEXT...