CardioTouch3000 Operation Manual
Base line drift is caused by the patient‘s breathing. It can be seen as ECG signals being
overlapped on a large parabola. The BASE LINE filter provides 0.05Hz, 0.1Hz and 0.2Hz, and
its set values will appear as Base:0.05Hz, Base:0.1Hz and Base:0.2Hz at the bottom of the
Output Form.
There are three types of an AC power noise filter; 60Hz, 50Hz and OFF. When set to
OFF, the filter will not be applied. If the filter is set to 50Hz or 60Hz, then the AC filter will
remove 50Hz or 60Hz power noise correspondingly. In the U.S., the AC filter should be
set to 60Hz. Check the properties of your power source before setting AC filter
properties.
The status of the AC filter can be checked on the bottom part of the printout -
indicating AC 60Hz, AC 50Hz or AC OFF