107 / 120
E3KOM-7.00
CardioTouch3000 Operation Manual
3) Trouble Shooting
Printing was not successful:
The cover may not be closed properly. Please close the cover and print again.
When using the battery power, there was a beep for more than 3 times and a
message came as shown below:
The battery is low. Please connect to the AC power to use.
The signal is printed out with noise:
First, check that you set up the AC power filter. If it still makes noise with the AC
power supply, please connect grounding line to the grounding electrode of
the system.
The grounding should not be used with the AC power‘s grounding and please
connect it to the metal around the patient‘s bed or the building.
When The external USB keyboard is not working properly:
Connect the external USB keyboard to the USB port before you turn on the
power of CardioTouch3000. If you connect the external USB keyboard to the
USB port after you turn on the power of CardioTouch3000, you should turn off
the power of CardioTouch3000 and turn on again.
If it is still not working, please contact our customer service center.
System messages
The following system messages may occur while you are operating this system.
You may be required to perform some action. If you perform the recommended
actions and the condition still remains, contact authorized service personnel.
Содержание CardioTouch3000
Страница 1: ...CardioTouch3000 Operation Manual Ver 7 00 2019 05 23 www ebionet com...
Страница 48: ...48 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 49: ...49 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 50: ...50 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 51: ...51 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 12 CH Record Report Form BEAT REPORT TEXT...