113 / 120
E3KOM-7.00
CardioTouch3000 Operation Manual
Guidance and manufacturer’s declaration – electromagnetic immunity
The CardioTouch3000 is intended for use in the electromagnetic environment specified below.
The customer or the user of the CardioTouch3000 should assure that it is used in such an
environment.
IMMUNITY
test
IEC 60601
test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
±
1 kV for
input/output lines
± 2 kV for power supply
lines
±
1 kV for input/output
lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips,
short
interruptions
and
voltage
variations
on power
supply
input lines
IEC 61000-4-
11
<5%UT
(>95%dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5, 6 cycles
70 %UT
(30 %dip in UT)
for 25,30 cycles
<5%UT
(>95%dip in UT)
for 5 s
<5%UT
(>95%dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5, 6 cycles
70 %UT
(30 %dip in UT)
for 25,30 cycles
<5%UT
(>95%dip in UT)
for 5 s
Mains power quality should
be that of a typical
commercial or hospital
environment.
If the user of the
CardioTouch3000 requires
continued operation during
power mains interruptions,
it is recommended that the
CardioTouch3000 be
powered from an
uninterruptible power supply
or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE
U
T
is the a.c. mains voltage prior to application of the test level.
Содержание CardioTouch3000
Страница 1: ...CardioTouch3000 Operation Manual Ver 7 00 2019 05 23 www ebionet com...
Страница 48: ...48 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 3CH Rhythm Report Form 6CH...
Страница 49: ...49 120 E3KOM 7 00 CardioTouch3000 Operation Manual Rhythm Report Form 12CH Record Report Form 3CH 1 RHY...
Страница 50: ...50 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 3 CH 3 RHY Record Report Form 6 CH 1 RHY...
Страница 51: ...51 120 E3KOM 7 00 CardioTouch3000 Operation Manual Record Report Form 12 CH Record Report Form BEAT REPORT TEXT...