[Filter Setup]
The signal output generated by the device may contain extra unwanted signals in
addition to the EKG wave such as AC noise, base line drift caused by respiration, noise
generated by an electrical field nearby, etc. Applying filters could help to reduce or
remove those unwanted signals from the output of the device.
To change filter information, enter the Filter Setup Menu either by moving the selection
block over and select it using the rotary key or by touching the menu on the touch screen.
Upon selection, the Filter Setup menu, shown below, will appear. Use either the rotary key
or touch screen to navigate and select menu items to edit. After editing, OK must be
selected in order to apply the changed value.