Chapter 3 Basic Operation
--22--
User Manual of Veterinary Monitor
3.1.3 Normal Operation Conditions
1.
Operating temperature: 0
℃~
40
℃
(32
℉~
104
℉
). (If the machine includes CO
2
module, the operating
temperature is 5
℃
~
40
℃
(41
℉
~
104
℉
)).
2.
Operating humidity:
15%
~
80%, non-refrigerated.
3. Atmospheric pressure: 442.5 mmHg
~
805.5 mmHg (59 kPa
~
107.4 kPa).
4.
Power source: AC 100 V
~
240 V, 50 Hz/60 Hz, frequency allowance ± 1Hz; DC 14.8 V, 4.4 Ah.
Attention
If the Monitor got condensation problem because of transfer from one place to another which lead to
temperature difference; or if the Monitor got moisture problem, you should wait until the
condensation or moisture problem disappears before using it again. If the Monitor works abnormal
still, stop using it and contact your dealer or our company immediately.
Warning
Please insure that the monitor is operated and stored in the required environment. Or, the technical
specifications described in the manual may not be reached, or unanticipated consequences such as
monitor damage may be caused.
3.2 Operation Preparation
3.2.1 A.C. Power Supply Connection
Please check the monitor and associated module states before A.C. power connection.
A.C. power connection procedures:
Be sure that the current A.C. power satisfies the following specifications: AC 100V
~
240V, 50 Hz/60 Hz.
Use the power cord supplied with the monitor; insert one end of the cord into the power interface on the
monitor, and the other end into a single-phase outlet with protective grounding.
Use the specialized grounding wire supplied with the monitor to connect the monitor to the protective ground
terminals.
Special attention: insure that the monitor has normal grounding.
Attention
In order to use battery power, the batteries have to be charged after monitor transportation or
storage. To turn on the monitor without an A.C. power connection, the monitor may not work
normally due to insufficient power supplied by the batteries.
This series of monitors are not suitable for connecting to CISPR11 provisions of public power.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......