Chapter 6 Parameter Monitoring
--82--
User Manual of Veterinary Monitor
module is not in use to prolong the service life of the water tank and module.
The filled dewatering bottles must be quickly replaced and exchanged to prevent any damage to the
module.
Don
’t use this module for expiratory gas measure when a dewatering bottle is not connected, for the
moist expiratory gases may cause errors to the measure and may shorten the module
’s service life
due to accumulation of humid gases.
6.8.3 Main-stream CO
2
Module
Warning
Do not use in the presence of flammable anesthetics or other flammable gasses. Use of the
CAPNOSTAT5 sensor in such environment may present an explosion hazard.
Electrical Shock Hazard: always disconnect the CAPNOSTAT5 sensor before cleaning. Do not use
if it appears to have been damaged. Refer servicing to qualified service personnel.
Do not position the sensor cables or tubing in any manner that may cause entanglement or
strangulation.
Reuse, disassembly, cleaning, disinfecting or sterilizing the single animal use CO
2
airway adapters
may compromise functionality and system performance leading to a user or animal hazard.
Performance is not guaranteed if an item labeled as single animal use is reused.
Inspect the CO
2
airway adapters for damage prior to use. Do not use the CO
2
airway adapters if
they appear to be damaged or broken.
Replace the CO
2
airway adapters if excessive secretions are observed.
If the CO
2
waveform (Capnogram) appears abnormal, inspect the CO
2
airway adapters and replace
if needed.
Monitor the CO
2
waveform (Capnogram) for elevated baseline. Elevated baseline can be caused by
sensor or animal problems.
Periodically check the CAPNOSTAT5 sensor and tubing for excessive moisture or secretion
buildup.
Do not operate the CAPNOSTAT5 sensor when it is wet or has exterior condensation.
Attention
Use only accessories provided by manufacturer.
Do not sterilize or immerse the CAPNOSTAT5 sensor in liquids.
Do not clean the CAPNOSTAT5 sensor and accessories except as directed in this manual.
It is recommended that the CO
2
sensor be removed from the circuit whenever an aerosolized
medication is delivered. This is due to the increased viscosity of the medications which may
contaminate the sensor windows, causing the sensor to fail prematurely.
Do not apply excessive tension to the CAPNOSTAT5 sensor cable.
This product and its accessories are latex free.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......