Append C Alarm Information
User Manual of Veterinary Monitor --151--
ECG
Alarm Information
Triggering Condition
Treatment Measure
ECG RLF C Lead Off
The connection between the electrode
and the animal is loose or fallen, or
the connection between lead line and
main cable is loosened. (Integrated
display is used for showing all
alarms, so that the user may easily
view all the information on a lead
off.).
Check the connection between electrode and
animal, as well as the connection between
lead line and main cable.
NIBP
Alarm Information
Triggering Condition
Treatment Measure
NIBP Measure Timeout
Failure occurs during measuring,
resulting in the system cannot make
an analysis and calculation.
Check the animal connection or change the
cuff. Then, restart the equipment to try again.
If the error remains, please contact the
manufacturer for repair.
NIBP Pressure Outrange
The animal‘s blood pressure exceeds
the measuring range.
Check the airway connection or change the
cuff. If the error remains, please contact the
manufacturer for repair.
NIBP Pressure Guard
The airway might be blocked.
Check the airway and measure again.
NIBP Arm Movement
The animal arm has moved.
Check the animal condition and stop the
animal from moving the arm.
NIBP Signal Too Weak
It might be that the animal‘s pulse is
too weak or the cuff is too loose.
Check the animal condition and put the cuff
to an appropriate position. If the error
remains, please change the cuff. If the
problem remains unsolved, please contact the
manufacturer for repair.
Cuff Type Error
The cuff is not compatible with the
setting of animal type.
Confirm the animal type or change the cuff.
Cuff Leakage
The NIBP cuff is not correctly
placed, or not properly connected, or
the airway has leakage.
Check the airway connection or change the
cuff. If the error remains, please contact the
manufacturer for repair.
Cuff Loose
The NIBP cuff is not correctly
placed, or not properly connected, or
the airway has leakage.
Check the air tube connection or change the
cuff. If the error remains, please contact the
manufacturer for repair.
Cuff
Enlaced
or
Air-Logged
The cuff or airway is blocked.
Check the airway and measure again. If the
error
remains,
please
contact
the
manufacturer for repair.
NIBP Measure Failed
During measuring, the system failed
and cannot make analysis.
Check the animal condition; and check the
connection or change the cuff. If the error
remains, please contact the manufacturer for
repair.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......