Append C Alarm Information
User Manual of Veterinary Monitor --149--
Appendix C Alarm Information
Physiological Alarm Information
Physiological Parameters
Alarm Information
Triggering Condition
Treatment Measure
xx Too High
xx value exceeds the alarm upper limit. Check the physiological condition of the
animal and confirm if the setting of animal
type and alarm limit is suitable to the
animal.
xx Too Low
xx exceeds the alarm lower limit.
Attention: xx represents the physiological parameter or name of a module, e.g. HR, ST-
Ⅰ
, SpO
2
, NIBP Systolic
Blood Pressure and RR, etc.
ECG
Alarm Information
Triggering Condition
Treatment Measure
ECG Signal Too Weak
The animal ECG signal is too weak.
Check the animal state, electrode and lead
cable.
Asystole
Heart beat NOT detected when preset
cardiac arrest threshold time has
passed
If the animal suffers arrhythmia, check the
animal state, electrode and lead cable.
Check if the setting of arrhythmia trigger
threshold is suitable to the animal.
VFib/VTac
Fibrillating waves last consistently for
6s//Dominant rhythm of the adjacent
ventricular heart beats (V) and the
heart rate is greater than the upper
limit of ventricular tachycardia
Extreme Tachycardia
Heart rate exceeds extreme tachycardia
threshold
Extreme Bradycardia
Heart
rate
lower
than
extreme
bradycardia threshold
Ventricular Rhythm
Lead
rhythm
of
the
adjacent
ventricular beat exceeding the number
of idioventricular rhythm threshold,
and the heart rate is lower than VT
rate.
Ventricular Bigeminy
Rhythm N, V, N and V
Ventricular Trigeminy
Rhythm N, N, V, N, N, V
Irregular Rhythm
Continuous irregular rhythm
PVCs/min
PVCs/min exceeds preset higher limit
Run PVCs
>
2
More than 2 continuous PVCs in the
last minute
Couplet PVCs
Paired PVCs detected in the last
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......