Chapter 6 Parameter Monitoring
--58--
User Manual of Veterinary Monitor
Manual Mode
In the Manual mode, users are required to set the respiratory detection level. Select [Upper Line] and
[Lower Line] to manually relocate the detection level dotted line on the respiratory waveforms. Once
determined, the detection level will not automatically adapt to different depths of respiration. Thus,
users are required to readjust the detection level to adapt to changes in the depths of respiration
according to actual conditions.
Attention
In case the respiratory waves are not displayed on the respiratory oxygenation diagram interface or
main screen, the detection level is still not adjustable even if the [Detect. Mode] is set Manual.
Respiration Alarm Parameter Setup
Select [Alarm Setup>>] to enter the menu [PAR.Alarm], then users can set RR alarm attributes, including:
Alarm Switch, Alarm Higher/Lower Limit, Alarm Level and Alarm Record On/Off.
6.3 PR
6.3.1 PR General Description
The mechanical movement of the heart causes pulsation of arteries, and the PR (pulse rate) can be obtained by
measuring this pulsation. The PR numerical value can be obtained through SpO
2
or any arterial pressure (IBP).
6.3.2 PR Source
Pulse is displayed in PR parameter region.
Figure 6.11 PR Parameter Region
When effective pulse origin is displayed in the PR parameter region, the pulse rate (PR) of the pulse can be:
Detected as the system pulse and an alarm triggered when users select PR as an alarm source.
Stored in the monitor database and reviewed in tendency diagram and tendency chart.
Transmitted to the central monitoring network when network is available.
PR Source Setup
Select the PR parameter region or open the [SpO
2
Setup] menu, and then select [PR Source]: Auto, SpO
2
, IBP1
and IBP2. When Auto is selected, the system selects an item from the selection list as its PR Source. The system
automatically switch the [PR Source] to [Auto] when the current PR source is not available.
6.3.3 Alarm Source Setup
As referred to Alarm Source Setup in section 6.1.6.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......