Chapter 6 Parameter Monitoring
User Manual of Veterinary Monitor --83--
After the life cycle of the CAPNOSTAT5 sensor and its accessories have been met, disposal should
be accomplished following national and local requirements.
Nitrous oxide, elevated levels of oxygen and helium can influence the CO
2
measurement. Please
setup gas compensation according to actual state.
Barometric pressure compensation is required to meet the stated accuracy of the CAPNOSTAT5
sensor.
Do not place the combined CO
2
sensor between the ET tube and the elbow (10~20 kg or
>
20 kg
circuit), as this may allow animal secretions to block the adapter windows.
Position the combined CO
2
sensor with its windows in a vertical and not a horizontal position: this
helps keep animal secretions from pooling on the windows.
Preparing to Measure CO
2
(Mainstream, CAPNOSTAT5)
1. Attaching the CAPNOSTAT 5 sensor cable.
To attach the CAPNOSTAT 5 sensor cable, plug the cable into CO
2
socket on the left panel of monitor by
matching the key on the cable to the key on the connector.
Attention
To remove the sensor cable from the monitor, grasp the collar surrounding the cable and pull
up.
2. Selecting a mainstream airway adapter.
Select an airway adapter based on the animal‘s size, ET tube diameter and monitoring situation. For more
information refer to the following or contact manufacturer.
Table 6.4 Airway Adapter Type
Airway Adapter Type
ET Tube Diameter
SPU*10
~
20 kg /
>
20 kg
>
4.0 mm
>
20 kg (Reusable)
>
4.0 mm
SPU*
<
10kg./0
~
20 kg
≤4.0 mm
<
10 kg. (Reusable)
≤4.0 mm
*SPU=Single Animal Use
3. Attaching the airway adapter to the CAPNOSTAT 5 sensor.
Before attaching the airway adapter to the CAPNOSTAT 5 sensor, verify that the airway adapter windows are
clean and dry. Clean or replace the adapter if necessary.
Follow these steps:
1) Align the arrow on the bottom of the airway adapter with the arrow on the bottom of the sensor.
2) Press the sensor and airway adapter together until they click.
3) Wait for the airway adapter and sensor to warm up.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......