Chapter 6 Parameter Monitoring
--70--
User Manual of Veterinary Monitor
[Stop All.] to terminate NIBP measure.
2.
Automatic Measure
Select [Measure Mode] and then set: Auto. Press the button [NIBP] on the front panel or [Start NIBP] to
manually start the first measure, and the system will automatically and repeatedly perform NIBP
measure according to the time intervals defined in [Interval] when the first measure is done.
If the button [NIBP] is pressed during the idle time between automatic measures, a manual measure will
be conducted. The monitor will continue automatic measures when the manual one is completed.
Press either the button [NIBP] on the front panel or [Stop NIBP] at any time the automatic measure is in
process, and the current NIBP measure will end, but the system will continue its automatic measure.
Press the button [Stop All.] in the window [NIBP Setup] at any time the automatic measure is in process,
all measuring tasks will be instantly ended.
3.
Continuous Measure
Select [Measure Mode] and then set: STAT. Press the button [NIBP] on the front panel or [Start NIBP]
to start continuous measure which occurs every 5 minutes.
Press the button [NIBP] on the front panel or [Stop NIBP] and [Stop All] at any time the continuous
measure is in process to terminate the continuous measure, and the monitor then returns to the Manual
Measure mode.
Attention: if the measure mode is set as continuous measure, when the monitor is turned off and then
boot, NIBP measure mode will be back to manual measure.
Warning
Please regularly check the color, warmth and sensitivity of the animal
’s distal limbs in the process of
measuring blood pressure under Auto and STAT modes. Once any abnormality is found, please
relocate the cuff or terminate blood pressure measure.
Interval Setup
Select [Interval] in the menu [NIBP Setup] when the measure mode is Auto, under the auto mode, the
monitor automatically and repeatedly conducts NIBP measures according to the preset time intervals.
Attention
This option is available only when the [Measure Mode] is Auto.
Alarm Setup
Select [Alarm Setup>>] in the menu [NIBP Setup] and then set in the pop-up menu the alarm attributes of
each NIBP option parameter.
Start /Stop NIBP Measure
Select this option to start or stop NIBP measure in accordance with the preset mode.
Stop All.
Select this option at any time when the measure is in process to terminate all measuring tasks.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......