Chapter 6 Parameter Monitoring
--66--
User Manual of Veterinary Monitor
2.
The subject should be in a horizontal position and the cuff should be placed level with the heart for the most
accurate results, other postures may cause inaccurate measures.
3.
Neither movement nor cuff touch is allowed before and during measure and the gas valve connecting the cuff
and the monitor should be guaranteed smooth and without being twisted.
4.
Too small measuring intervals are not suitable (optimally longer than 2min), for the excessively short
intervals between continuous measures can cause extrusion of arms, resulting in decreased blood flows and
then decreased blood pressure.
5.
It should be judged by considering clinical evaluations to determine whether or not automatic blood pressure
measure is adapted for animals with severe blood coagulation mechanism disorders, for the friction between
limbs and the cuff may cause risks for haematoma.
6.
The cuff inflation pressure is automatically regulated by computer based on the last measured values in the
successive blood pressure measure. It is possible that the sphygmomanometer is unable to give any result
after the first inflation when the blood pressure is elevated or the animal is changed; this series of monitors
will automatically regulate inflations and continue measuring until there are results produced, but the
measure should be conducted no more than 4 times.
7.
It should be guaranteed that correct animal categories are selected while measuring animals from 10~20 kg
and animals < 10 kg. Improper selections will possibly put animals into dangerous conditions, for relatively
high-pressure animals > 20 kg setting is not suitable for animals from 10~20 kg and animals <10 kg.
8.
It is possible that no measure results will be produced when selections for animals from 10~20 kg or animals
< 10 kg are employed for animals > 20 kg measure, even though no harm will be inflicted.
9.
The limbs used by the cuff will be possibly accompanied with purpura, ischemia and neurologic damage due
to drawn-out non-invasive blood pressure measure when the Auto mode is used. Please check the color,
warmth and sensitivity of animal‘s distal limbs regularly while monitoring. Once any abnormality is found,
please relocate the cuff or terminate the blood pressure measure.
10.
If the animal has irregular heartbeat causing by arrhythmia, the measurement will not reliable, and it will cost
more time to measure the blood pressure.
Warning
NIBP measures are not allowed for animals with sickle cell disorders and present or possibility of
incurring skin damage.
It should be judged by considering clinical conditions to determine whether or not automatic blood
pressure measure is adapted to animals with severe thrombi disorders, for the limbs used for the
cuff have a possibly risk of haematoma.
Cuff are not allowed on limbs with venous transfusion or cannula, for the transfusion is lowered or
blocked due to the inflated cuff, which will possibly lead to tissue damage around the cannula.
Please use other procedures to check animal
’s vital signs and then check whether or not the
monitor is working well when the accuracy of the measured results is open to suspicion.
The Monitor is suitable for used with electrosurgical units (ESU). To reduce the hazard of burns
during high-frequency surgical procedure, ensure that the monitor
’s cables never come into contact
with the high-frequency surgical units.
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......