Append B Product Specifications
User Manual of Veterinary Monitor --143--
Inflation
Pressure
Setting
10
~
20 kg: 80 mmHg
~
210 mmHg (10.7 kPa
~
27.9 kPa)
<
10 kg: 60 mmHg
~
140 mmHg (8.0 kPa
~
18.6 kPa)
Default
of
Initial
Inflation Pressure
>
20 kg: 160 mmHg (21.3 kPa)
10
~
20 kg: 140 mmHg (18.6 kPa)
<
10 kg: 90 mmHg (12.0 kPa)
Measuring Tolerance of
Pressure
Source
Testing
±3 mmHg (±0.4 kPa)
Overpressure
Protection
>
20 kg state: When the pressure in cuff exceeds 297 mmHg (39.5 kPa)
±
3 mmHg
(0.4 kPa), the control valve shall relieve the pressure.
10
~
20 kg state: When the pressure in cuff exceeds 240 mmHg (31.9 kPa)
±
3 mmHg
(0.4 kPa), the control valve shall relieve the pressure.
<
10 kg state: When the pressure in cuff exceeds 147 mmHg (19.6 kPa)
±
3 mmHg
(0.4 kPa), the control valve shall relieve the pressure.
Alarm
Limit
Specifications
Range
Upper Limit of Systolic
Blood Pressure
>
20 kg: (Lower limit+5) mmHg
~
270 mmHg ( (Lower limit+0.7) kPa
~
35.9 kPa)
10
~
20 kg: (Lower limit+5) mmHg
~
200 mmHg ( (Lower limit+0.7) kPa
~
26.6 kPa)
<
10 kg: (Lower limit+5) mmHg
~
135 mmHg ( (Lower limit+0.7) kPa
~
18.0 kPa)
Lower
Limit
of
Systolic Blood Pressure
>
20 kg: 41 mmHg
~
(Upper limit-5) mmHg (5.3 kPa
~
(
Upper limit
-
0.7
)
kPa)
10
~
20 kg: 40 mmHg
~
(Upper limit-5) mmHg (5.3 kPa
~
(
Upper limit-0.7
)
kPa)
<
10 kg: 40 mmHg
~
(Upper limit-5) mmHg (5.3 kPa
~
(Upper limit-0.7
)
kPa)
Upper Limit of Mean
Blood Pressure
>
20 kg: (Lower limit+5) mmHg
~
230 mmHg ( (Lower limit+0.7) kPa
~
30.6 kPa)
10
~
20 kg: (Lower limit+5) mmHg
~
165
mmHg ( (Lower limit+0.7) kPa
~
21.9.0
kPa)
<
10 kg: (Lower limit+5) mmHg
~
110 mmHg ( (Lower limit+0.7) kPa
~
14.6
kPa)
Lower Limit of Mean
Blood Pressure
>
20 kg: 20 mmHg
~
(Upper limit-5) mmHg (2.7 kPa
~
(
Upper limit-0.7) kPa)
10
~
20 kg: 20 mmHg
~
(Upper limit-5) mmHg (2.7 kPa
~
(
Upper limit-0.7
)
kPa)
<
10 kg: 20 mmHg
~
(Upper limit-5) mmHg (2.7 kPa
~
(Upper limit-0.7) kPa)
Upper
Limit
of
Diastolic
Blood
Pressure
>
20 kg: (Lower limit+5) mmHg
~
210
mmHg ( (Lower limit+0.7) kPa
~
27.9 kPa)
10
~
20 kg: (Lower limit+5) mmHg
~
150
mmHg ( (Lower limit+0.7) kPa
~
20.0 kPa)
<
10 kgv: (Lower limit+5) mmHg
~
100
mmHg ( (Lower limit+0.7) kPa
~
13.3 kPa)
Lower
Limit
of
Diastolic
Blood
Pressure
>
20 kg: 11 mmHg
~
(Upper limit-5) mmHg (1.4 kPa
~
(Upper limit-0.7
)
kPa)
10
~
20 kg: 11 mmHg
~
(Upper limit-5) mmHg (1.4 kPa
~
(Upper limit-0.7) kPa)
<
10 kg: 10 mmHg
~
(Upper limit-5) mmHg (1.3 kPa
~
(
Upper limit-0.7) kPa)
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......