Append C Alarm Information
--152--
User Manual of Veterinary Monitor
IBP System Error
Air pump, A / D sampling or pressure
sensor error, or pointer error in the
software running, or system needs to
be calibrated.
Please contact the manufacturer for repair.
SpO
2
Alarm Information
Triggering Condition
Treatment Measure
SpO
2
Sensor Off
The sensor is fallen from the animal
or module, resulting in error.
Check the sensor connection.
SpO
2
Sensor Disconnect
SpO
2
probe is not properly connected. Check the connection of SpO
2
probe.
Resp
Alarm Information
Triggering Condition
Treatment Measure
Resp Module Interrupt
The module circuit is interfered with
Restart the equipment. If the error remains,
please contact the manufacturer for repair.
Temp
Alarm Information
Triggering Condition
Treatment Measure
T Module Disconnect
The
temperature
probe
is
not
correctly connected or it is damaged.
Check the temperature probe and its
connection.
The
measured
values
beyond the measurement
range
The Temperature measurement result
is not within the ranging of 0
℃
~
50
℃
.
Check if the type of temperature probe is
consistent with the setting of the monitor.
Check if the temperature probe is not well
connected or damaged.
CO
2
Alarm Information
Triggering Condition
Treatment Measure
CO
2
Sensor Off
The CO
2
sensor is not correctly
connected.
Confirm that the CO
2
sensor has been
correctly connected.
CO
2
Sensor Too Hot
The CO
2
sensor temperature is too
high.
Check and stop using or change the sensor.
CO
2
Sensor Too Cold
The CO
2
sensor temperature is too
low
CO
2
Pressure Too High
The pressure of airway is abnormal
Check the animal and airway connection.
Then restart the monitor.
CO
2
Pressure Too Low
CO
2
Airpressure Too High
The environment where the monitor
is located affects the pressure.
Check the airway connection and confirm if
the environment conforms to the monitor
specifications and if there is any special factor
affecting the environment pressure.
CO
2
Airpressure Too Low
CO
2
Gascircuit Jam
The airway is blocked
Check the airway and eliminate the blocking.
CO
2
Basin Off
The
water
bath
is
improperly
connected.
Check the water bath connection.
CO
2
Zero Error
The airway is improperly connected. Check the airway connection. Make zero
Содержание PM-900
Страница 2: ......
Страница 12: ...Content VIII User Manual of Veterinary Monitor The Blank Page ...
Страница 14: ...Preface 2 User Manual of Veterinary Monitor The Blank Page ...
Страница 42: ...Chapter 3 Basic Operation 30 User Manual of Veterinary Monitor The Blank Page ...
Страница 132: ...Chapter 12 Battery 120 User Manual of Veterinary Monitor The Blank Page ...
Страница 142: ...Chapter 15 Troubleshooting and Solutions 130 User Manual of Veterinary Monitor The Blank Page ...
Страница 176: ...Append D Factory Default Setup 164 User Manual of Veterinary Monitor The Blank Page ...
Страница 181: ......