Append C Alarm Information
--154--
User Manual of Veterinary Monitor
equipment.
Head of Print Hot
The recorder has worked too long
Clear the print task and output the records
after the machine has cooled.
Voltage of Battery Too
Low
The battery voltage is low and cannot
maintain long-time monitoring
Switch to AC power supply. Power supply by
battery can only be used when the battery is
fully recharged.
Very Low Voltage, Shortly
Logout
The battery voltage is too low. To
avoid data loss due to low power, the
system
will
soon
activate
the
automatic shutdown procedure.
Switch to AC power supply. Power supply by
battery can only be used when the battery is
fully recharged.
Attention
When different levels of alarms exist together, the alarm sound of the highest level will be heard.
Under
‘Alarm Pause’ state, the monitor will not process any alarm information.