MERCURY PLUS
1-239
the board
connection
responded to a
command.
remote control
board (serial cable)
and make sure that
it is being correctly
powered in the
panel.
1114-1202
Serious errors
inside the software.
Contact the
software office.
2001
Weighing target too
far in advance
This problem is
due to the incorrect
position of the
static weighing
target. Typically, a
static weighing
target that is too far
in advance makes
the piece to be
weighed stop after
having passed
beyond the end of
the belt as well.
Alternatively, if the
target is too far
back, the piece to
be weighed stops
too soon and the
previous piece will
have not yet left
the scale.
Proceed in the
following way to
modify the
"weighing target"
value:
1- access the
process again
2- press "MODIFY
DATA"
immediately,
before a package
has time to transit
3- modify the
"WEIGHING POS."
value
4- save the
changes and
access the
processing phase
again
This changes the
value of that plu.
If the problem still
persists even with
the minimum (or
maximum) value,
check the same
value entered in
the general
settings of the
machine.
2002
Label delayed
The piece has
reached the
application target
Possible causes:
- belt targets too far
in advance
Содержание Cigiemme MERCURY PLUS
Страница 1: ...MERCURY PLUS Instruction manual Code 81320114 EDITION 02 05 2016 January 2007 ...
Страница 2: ......
Страница 12: ...MERCURY PLUS 1 12 ...
Страница 19: ...1 19 MERCURY PLUS 1 4 10 Glossary Gives the meanings of the technical terms used in this manual ...
Страница 62: ...MERCURY PLUS 1 62 ...
Страница 94: ...MERCURY PLUS 1 94 5 9 List of menus Figure 5 2 1 3 menu list citi6117 en jpg ...
Страница 95: ...MERCURY PLUS 1 95 Figure 5 3 2 3 menu list citi6118 en jpg ...
Страница 96: ...MERCURY PLUS 1 96 Figure 5 4 3 3 menu list citi6119 en jpg ...
Страница 194: ...MERCURY PLUS 1 194 ...
Страница 196: ...MERCURY PLUS 1 196 Figure 6 1 Permit combinations citi6134 jpg ...
Страница 234: ...MERCURY PLUS 1 234 ...
Страница 250: ...MERCURY PLUS 1 250 ...
Страница 259: ......