5-3
Care and Maintenance
Tube Breakage
If a glass tube breaks, and all the glass is not contained in the bucket
or rotor, be sure to thoroughly clean the chamber.
WARNING
!
Be careful when examining or cleaning the
chamber and chamber gasket, as sharp glass
fragments may be embedded in their
surfaces.
• Examine the chamber gasket to make sure that no glass particles
are retained in it. Carefully remove any glass particles that may
remain.
• Carefully wipe away any glass particles that remain in the
chamber.
DECONTAMINATION
If the instrument and/or accessories are contaminated with radioactive
or pathogenic solutions, perform appropriate decontamination proce-
dures. Refer to Appendix A in
Rotors and Tubes
or publication
IN-175 to be sure the decontamination method will not damage any
part of the centrifuge.
STERILIZATION AND DISINFECTION
The centrifuge is finished with urethane paint. Ethanol (70%)
2
may
be used on this surface. See Appendix A in
Rotors and Tubes
or
publication IN-175 for chemical resistances of centrifuge and acces-
sory materials.
While Beckman Coulter has tested these methods and found that they
do not damage the centrifuge, no guarantee of sterility or disinfection
is expressed or implied. When sterilization or disinfection is a
concern, consult your laboratory safety officer regarding proper
methods to use.
2
Flammability hazard. Do not use in or near operating centrifuges.