![Beckman Coulter Avanti J-25 Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/beckman-coulter/avanti-j-25/avanti-j-25_instruction-manual_2719702062.webp)
4-2
Troubleshooting
• the diagnostic number and message,
• the operating situation when the diagnostic condition occurred
(such as rotor in use, speed, or load type), and
• any unusual environmental and/or operating conditions (such
as ambient temperature or voltage fluctuations).
➠
NOTE
The information provided in Table 4-1 is a user
guide, and is not a comprehensive checklist.
Table 4-1. Diagnostic Message Chart
*
In the U.S. call 1-800-742-2345. Outside the United States, contact your local Beckman Coulter office.
Diagnostic Number/
Message
Problem
Result
Recommended Action
P1/
Power failure, run
continued
Momentary power
failure: rotor does not
come to a complete stop
Run continues when
power resumes
Press
[
CE
]
to clear message.
P2/
Power failure, run
restarted
Power failure: rotor
comes to a complete
stop
Run will restart auto-
matically when power
resumes
Press
[
CE
]
to clear message.
L2/
Door not latched,
reclose door
L5/
Door not latched, call
Beckman service
Door not closed firmly
enough, or debris in
latch area may be
preventing latches from
closing securely
L2 error message
appears. If you reclose
the door repeatedly and
the problem continues,
the message will
change to say “L5 Door
not latched, call
Beckman Coulter
service”
Reclose the door firmly, placing your
hand over the drawing of the hand on
the door label. (As you close the door
your hand should be half-way between
the two latches, which is approxi-
mately 45° to the right of center front.)
Gently clean the latch areas with a dry,
lintless cloth. Be careful not to damage
sensitive electronics in the area. If L2
continues to appear after cleaning, call
Beckman Coulter Field Service.*
L3/
Door problem, refer
to manual
L9/
Door problem, refer
to manual
Foot pedal was
depressed before end
of run
Latches are not
operating properly
Press
[
CE
]
and try to restart the run.
If this does not work, call Beckman
Coulter Field Service.
Continued —