![Beckman Coulter Avanti J-25 Скачать руководство пользователя страница 49](http://html1.mh-extra.com/html/beckman-coulter/avanti-j-25/avanti-j-25_instruction-manual_2719702049.webp)
3-10
Operation
Action
Result
3. Press the next function key or
press [
ENTER
], then [
START
].
After you start the run, the HOLD indicator illuminates in the
TIME
display. The elapsed run time is displayed. The run continues until
[
STOP
] is pressed.
ENTERING RUN TEMPERATURE
Run temperature can be set from –20 to +40°C. If no value is entered,
the last-entered temperature is used.
Action
Result
1. Press [
TEMP
].
The
Temp:
field flashes.
2. Use the keypad to enter the
required temperature.
Entered temperature appears in the
Temp:
field.
3. Press the [
TEMP
] key again
to move the cursor to the
Max Temp:
field.
Entered temperature plus 4°C appears in the
Max Temp:
field, and
YES and NO soft keys appear. The default maximum temperature is
the set temperature plus 4°C. You can specify a different maximum
temperature, from +2°C above the set temperature to 44°C. You can
also specify whether or not you want the system to shut down if the
maximum temperature is reached. The centrifuge will always shut
down if the system temperature reaches 45°C.
4. Use the keypad to enter a
different maximum tempera-
ture, or accept the default.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Rotor : JA-25.50
Speed : 25 000 RPM = x G
Time : 00:00
Temp: - - Max Temp : - - °C
Accel : MAX Decel : MAX