3-18
Operation
Action
Result
1. Turn the key to the ZONAL
position.
Open-door operation up to 3000 rpm is enabled. JCF-Z appears in the
Rotor:
field.
2. Install the rotor. Leave the
door open when installation
is complete.
3. Press [
ROTOR
]. Press the
JCF-Z soft key. Enter the
remaining run parameters.
4. Press [
ENTER
], then [
START
].
The rotor begins accelerating to 2000 rpm, which is the default
loading speed. To select a different loading speed, enter a speed
of 3000 rpm or less before or after the rotor starts to spin.
If an
ω
2
t
run is specified,
ω
2
t
starts accumulating.
5. Load the rotor.
6. Close the door.
The rotor accelerates to the set speed.
If you entered a loading speed
other than 2000 rpm, re-enter the run speed after you close the door.
7. End the run.
When the run is complete (the run time has expired, the set
ω
2
t
value
has been reached, or [
STOP
] has been pressed), the rotor decelerates to
the default unloading speed of 2000 rpm. If required, you may enter a
different unloading speed of 3000 rpm or below after the rotor
reaches 2000 rpm.
ZONAL
Accel : MAX Decel : MAX
Rotor : JCF-Z
Speed : 0 RPM = x G
Time : 00:00
Temp : 25°C Max Temp : 29°C
JCF-Z
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Accel : MAX Decel : MAX
Rotor : JCF-Z
Speed : 20 000 RPM = x G
Time : 00:45
Temp : 20°C Max Temp : 24°C
SPEED
TIME
TEMP
A/D
OPT
ROTOR
JA
JS
JLA
JV
JE
JCF-Z