4-7
Troubleshooting
Action
Result
4. Lift the front panel up and off
the centrifuge; set it aside.
An inner front panel extends about halfway down from the centrifuge
top (see Figure 4-2). Below this inner panel is a black manual door
release interlock switch (see Figure 4-2).
Figure 4-2. Manual Door Release
5. Pull the interlock lever out
and to the left (at about a
45-degree angle), and while
holding it out, step on the
foot pedal.
Depending on the level of vacuum in the chamber at the time of the
power failure, the door may or may not open. If it opens, go to step 6.
If it doesn’t open, you will need to manually vent the vacuum. Go to
step 7.
6. After the door opens, first
release the foot pedal and
then release the interlock
lever.
Sample can be removed. Go to step 10.
START
ROTOR ID
SPEED
TIME
RPM
TEMP
RPM/RCF
ROTOR
SPEED
TIME
TEMP
A/D
OPT
ROTOR: JA-20
RPM
RCF
TIME: 00:00
TEMP: 25°C
MAX TEMP : 29°C SHUTDOW
N? YES
ACCEL: MAX
DECEL: MAX
HR:MIN
°C
CE
3
6
9
2
+
5
8
1
0
4
7
E
N
T
E
R
STOP
.
Manual Door Release
Interlock Lever
Retaining
Strip
®
Inner Front
Panel