3-9
Operation
Action
Result
4. Press the next function key or
press [
ENTER
], then [
START
].
The centrifuge begins calculating and displaying the accumulated
centrifugal effect.
Deceleration begins when the set
ω
2
t
value is reached. During decel-
eration, the red indicator light on the [
STOP
] key flashes. The
ω
2
t
continues to accumulate until the rotor stops spinning, so set and
actual
ω
2
t
values may be different at the end of the run.
Entering a Continuous (HOLD) Run
Action
Result
1. Press [
TIME
].
The HH:MM, W2T, and HOLD soft keys appear.
2. Press the HOLD soft key
to select the hold mode.
HOLD appears in the
Time:
field.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
ROTOR ID
SPEED
RPM/RCF
TIME
°C
TEMP
Rotor : JA-25.50
Speed : 25,000 RPM = x G
Time : 1.05E03
ω
2t
Temp : 25°C Max Temp : 29°C
HH:MM
W2T
HOLD
Accel : MAX Decel : MAX
EXP
06
rad2/sec
ROTOR ID
SPEED
RPM/RCF
TIME
HR:MIN
TEMP
Rotor : JA-25.50
Speed : 25 000 RPM = 0 x G
Time : 00:00
Temp : 25°C Max Temp : 29°C
HH:MM
W2T
HOLD
HOLD
Accel : MAX Decel : MAX
SPEED
TIME
TEMP
A/D
OPT
ROTOR
°C