![Beckman Coulter Avanti J-25 Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/beckman-coulter/avanti-j-25/avanti-j-25_instruction-manual_2719702041.webp)
3-2
Operation
SUMMARY OF RUN PROCEDURES
For runs at other than room temperature, precool or prewarm the rotor to the required temperature before
the run.
MANUAL RUN
1
Press the POWER switch to on (I).
2
Depress the foot pedal to open the chamber door.
3
Install the rotor according to instructions in the applicable rotor manual, then firmly
close the chamber door.
4
Press
[
ROTOR
]
, press the soft key to select rotor type, press the soft key to select the
rotor name.
5
Press
[
SPEED
]
, press the soft key to select RPM or RCF mode, then use the keypad to
enter the run speed (0 to 25 000 rpm).
6
Press
[
TIME
]
, press the soft key to select a time mode (HH:MM, Hold, or
ω
2
t
),
then use the keypad to enter the run time or
ω
2
t
value. (No entry is made in Hold
mode.)
7
Press
[
TEMP
]
, then use the keypad to enter the required run temperature.
8
Press [
A/D
] and press the MAX, SLOW, or TIME soft key. Or, skip this step and accept
the default acceleration rate (SLOW for the JS-24.38 and JS-24.15 rotors; MAX for all
other rotors).
9
Press [
A/D
] and press the MAX, SLOW, TIME, or OFF soft key. (Or, skip this step and
accept the default deceleration rate (SLOW for the JS-24.38 and JS-24.15 rotors; MAX
for all other rotors).
10
Check that all parameters are correct and that the door is closed. Press
[
ENTER
]
, then
press
[
START
]
.
11
Wait for the set time to count down to zero, or end the run by pressing
[
STOP
]
.
12
When the rotor stops, depress the foot pedal to open the chamber door.