![Beckman Coulter Avanti J-25 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/beckman-coulter/avanti-j-25/avanti-j-25_instruction-manual_2719702017.webp)
1-4
Specifications and Preinstallation Requirements
a
Maximum speeds are based on a solution density of 1.2 g/mL with the following exceptions: the JA-18.1 rotor is rated for a
density of 1.4 g/mL; the JCF-Z rotor is rated for a density of 1.45 g/mL; and the JE-6B rotor is rated for a density of 3 g/mL.
b
Maximum speed for the JA-30.50 Ti rotor is 25 000 rpm in the Avanti J-25I centrifuge, and 30 000 rpm in the Avanti J-30I centrifuge.
c
Maximum speed for the JS-24.38 and JS-24.15 rotors is 10 000 rpm in the Avanti J-25I centrifuge, and 24 000 rpm in the
Avanti J-30I centrifuge.
Rotor Profile and Description
Max
RPM
a
Max
RCF
(
×
g
)
Max
Capacity
Rotor
Manual
Number
JA-20 Fixed Angle, 34° (8 place)
r
max
= 108 mm
20 000
48 400
8
×
50 mL
J-TB-003
JA-18.1 Fixed Angle (24 place)
45° adapter
r
max
= 116 mm
25° adapter
r
max
= 112 mm
18 000
17 000
42 100
36 300
24
×
1.8 mL
24
×
1.8 mL
J-TB-037
JA-18 Fixed Angle, 23° (10 place)
r
max
= 132 mm
18 000
47 900
10
×
100 mL
J-TB-035
JA-17 Fixed Angle, 25° (6 place)
r
max
= 132 mm
17 000
39 800
14
×
50 mL
J-TB-017
JLA-16.250 Fixed Angle, 25° (6 place)
r
max
= 134 mm
16 000
38 400
6
×
250 mL
J-TB-072
JA-14 Fixed Angle, 25° (6 place)
r
max
= 137 mm
14 000
30 100
6
×
250 mL
J-TB-004