3-21
Operation
Action
Result
7. When the rotor speed has
stabilized, look through the
port in the centrifuge door
and turn the
DELAY
knob
until the elutriation chamber
in the rotor is synchronized
with the strobe. (The
chamber will appear to
be motionless).
Sample can now be injected into the lines. See the rotor instruction
manual for detailed instructions.
➠
NOTE
At the end of the run, after disassembling the
liquid and power lines, be sure to replace the
black rubber stoppers in the ports at the left side
of the centrifuge door. Failure to seal these
openings will prevent proper operation of the
FRS system, which will cause a diagnostic shut
down during the next run.
DELAY
Delay
Control
Knob
ELUTRIATOR