Troubleshooting
Table 4-1. Diagnostic Message Chart (continued)
* In the U.S. call 1-800-742-2345. Outside the United States, contact your local Beckman Coulter office.
†
Friction Reduction System.
Diagnostic Number/
Message
Problem
Result
Recommended Action
L4/
Door problem, refer
to manual
Door stays locked after
rotor stops spinning
Sample cannot be
retrieved
Restart the centrifuge and perform
a brief run. If door will still not
open after decel, call Beckman
Coulter Field Service.
*
C2/
Rotor over temp, run
stopped
Rotor temperature
exceeds maximum
temperature setting
Run shuts down with
maximum brake
Call Beckman Coulter Field
Service.
C3/
No temp control,
refer to manual
Refrigeration system
malfunction
Run shuts down with
maximum brake
• Check the air filter and replace if
dirty (see Section 5, Care and
Maintenance).
• Call Beckman Coulter Field
Service.
D1-5/
Drive error, call
Beckman Field
Service
Drive system problem
(for example, over
current, over/under
voltage, components
over temperature)
Run stops, usually with
no brake. Door may not
unlock for up to an hour.
Call Beckman Coulter Field
Service. Before trying to open the
door, listen carefully and make
sure that no sound is coming from
the chamber (indicating a spinning
rotor). Follow the directions under
ACCESSING THE ROTOR IN CASE
OF POWER FAILURE
, below.
Fl, F4, F5/
FRS
†
failure, refer to
manual
Required FRS level not
reached
Run shuts down with
maximum brake
• Check and clean door sealing
area and door gasket.
• Wipe excess moisture from
chamber.
• Make sure that elutriator port
stoppers are in place at left side
of door seal.
F2/
FRS failure, refer to
manual
FRS level too high
Run continues
Press
[
CE
]
. If F2 message
reappears, perform same actions
as for Fl error.
F3/
FRS failure, refer to
manual
FRS level too high
Run shuts down with
maximum brake
Call Beckman Coulter Field
Service.
S1/
System error
There is a problem with
the system control soft-
ware, EPROM, or RAM
System shuts down
Call Beckman Coulter Field
Service.
Continued —