![Beckman Coulter Avanti J-25 Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/beckman-coulter/avanti-j-25/avanti-j-25_instruction-manual_2719702055.webp)
3-16
Operation
Action
Result
7. Press the SAVE soft key
when programming is
complete.
The program is saved.
8. Press the EXIT soft key to
exit program mode.
MODIFYING A PROGRAM
Action
Result
1. Press the [
OPT
] key.
The
Prog Recall:
field flashes, and the message “Enter program
number (1–30)” appears.
2. Press [
OPT
] again to move the
cursor to the
Setup:
field.
Enter the number of the
program to be modified.
The program number appears in the
Setup:
field. A message appears
asking if you want to overwrite the program. YES and NO soft keys
appear.
3. Press the YES soft key.
The step 1 parameters appear.
4. To see step 2 parameters,
press the STEP2 soft key.
The step 2 parameters appear.
5. Modify the parameters for
either step. Press [
ENTER
]
when done. Repeat for the
other step, if required.
The newly entered parameters appear.
SPEED
TIME
TEMP
A/D
OPT
ROTOR
Accel : MAX Decel : MAX
Prog Recall: - - Setup: 1 Step #
YES
NO
Rotor : JA-20
Speed : 0 RPM = x G
Time : 00:00
Temp : 25°C Max Temp : 29°C Shutdown? No
Program number used, overwrite Y/N?