07-19-B1-729
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
5-3
Configuration Option Feature
Modifying the Configuration Option Settings
2. Occlusion Alarm Level: The increase in pressure required to trigger a
downstream occlusion alarm.
1: LEVEL 1 (3.0 to 13.0 psi,
nominal 7 psi)
2: LEVEL 2 (8.0 to 18.0 psi,
nominal 12 psi)
3: LEVEL 3 (13.0 to
23.0 psi, nominal 17 psi)
LEVEL 1
3. Audible Switchover: Determines whether or not an audible alert tone
is generated when either pump switches from the secondary program to
the primary program.
1: OFF
2: ON
OFF
4. Number of Automatic Restarts: Determines whether or not the pump
automatically restarts after a downstream occlusion, and if so, how
many restarts can occur before the pump remains in alarm. When this
parameter is set to anything other than 0, it is enabled. The selected
number corresponds to the number of automatic restarts. When set to 0,
the feature is disabled.
0 - 9
3
5. Door Open Required: Determines when the pump door must be
opened after a downstream occlusion alarm. When the door is not
opened and the pump is started within two minutes of the alarm, the
pump uses the pressure at which the alarm occurred as the baseline for
the next alarm. You may wish to set this option to force the user to open
the pump door. This action resets the baseline and encourages the user
to relieve the downstream pressure, thereby lowering the alarm
threshold.
Example
: Suppose the initial pressure is approximately 1 psi and the
occlusion alarm is set to LEVEL 1 or approximately 7 psi. The first
alarm occurs at approximately 1+7 or 8 psi. This value represents the
baseline pressure that is used to calculate the next alarm unless the door
is opened, the downstream pressure is relieved, and a new baseline is
set.
1: OFF
2: ON
OFF
6. Air Bubble Alarm Size: Determines the air bubble size that causes an
AIR alarm.
1: MIN (average 50
μ
L)
2: NORM (average 75
μ
L)
NORM
7. Alarm Off Interval: Selected setting is equivalent to the number of
seconds between each occurrence of the three-beep alarm tone.
1 - 7
1
8. Alert Off Interval: Selected setting is equivalent to the number of
seconds between each occurrence of the one-beep alert tone.
1 - 7
7
9. Maximum Rate of Infusion: Sets the maximum programmable
primary infusion rate of both pumps. When the maximum primary rate
is 999 or higher, the maximum secondary rate is 999. At values below
999, the maximum secondary rate matches the value of the maximum
primary rate.
1 - 1999 mL/hr
1999 mL/hr
10. Maximum VTBI: The maximum volume that either pump can be
programmed to infuse.
1 - 9999 mL
9999 mL
Table 5-1
Configurable Settings — continued
Parameter Description
Setting Options
Factory
Settings
Содержание Flo-Gard 6301
Страница 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 20: ...1 General Information 1 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 30: ...2 Controls and Indicators 2 10 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 34: ...3 Administration Sets 3 4 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 36: ...4 Nurse Call 4 2 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 74: ...7 Troubleshooting Chart 7 28 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 110: ...8 Disassembly Reassembly 8 36 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 200: ...13 Overview 13 2 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 201: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 3 Figure 13 1 System Block Diagram ...
Страница 205: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 7 Figure 13 5 Power Supply Section ...
Страница 206: ...13 8 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 6 DC Power Supply ...
Страница 210: ...13 12 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 10 Master CPU Schematic ...
Страница 213: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 15 Figure 13 13 Address Decoder ...
Страница 216: ...13 18 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 16 Slave CPU Schematic ...
Страница 220: ...13 22 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 20 CPU PCB Sensor Block ...
Страница 221: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 23 Figure 13 21 PPI Block 1 ...
Страница 222: ...13 24 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 22 PPI Block 2 ...
Страница 224: ...13 26 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 24 Air Sensing Block ...
Страница 226: ...13 28 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 26 Display PCB ...
Страница 229: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 31 Figure 13 29 LCDs for Pump 1 ...
Страница 230: ...13 32 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 30 LCDs for Pump 2 ...
Страница 231: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 33 Figure 13 31 LED Lamps ...
Страница 232: ...13 34 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 32 Audible Alarm PCB ...
Страница 233: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 35 Figure 13 33 Backup Buzzer PCB ...
Страница 234: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 36 Figure 13 34 Terminal PCB ...
Страница 235: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 37 Figure 13 35 Accessories ...
Страница 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 251: ...07 19 B1 729 ...