9
Upstream Occlusion Sensor Calibration
9-10
FLO-GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL
07-19-B1-729
4.
Perform the Operational Checkout procedures in Chapter 10,
“Operational Checkout”.
Upstream Occlusion Sensor Calibration
1.
Connect the AC plug of the pump to a 115 ±5.0 VAC 60 Hz power
source.
2.
Enter Automatic Test Mode 1 and check the calibration as described in
“Automatic Test Mode 1: Calibration Mode 1” on page 9-14, steps 7
through 5.
Q
If the value is within specification, proceed to the following step.
Q
If the value is out of specification, calibrate the sensor using the
following procedure.
2.1
Remove the pump head door cover. See “Replacement of Pump
Head Door Cover” on page 8-9.
2.2
Loosen and remove the occlusion sensor stop (Figure 11-5 and
Figure 11-10 , item 19).
Note:
If it is stuck, apply a small amount of isopropyl alcohol to the threads.
W
ipe
the threads clean.
2.3
Lay the pump on its back and place the thickness gauge on the
upstream occlusion sensor, then close the pump door.
2.4
Apply Loctite 211 or 222 to the threads of the occlusion stop
and reassemble the stop in the door.
2.5
Slowly turn the occlusion stop with a screwdriver until the
value in the PRI RATE window of the Main Display is between
3273 and 3277.
2.6
Knock on the center of the door by the back of fingers and
check that the reading does not change more than 2 counts.
2.7
Allow the Loctite 211 or 222 to set.
2.8
Remove the thickness gauge and stand the pump upright.
2.9
Install the pump head door cover. See “Replacement of Pump
Head Door Cover” on page 8-9.
2.10
Turn off the pump.
3.
The following procedure confirms that the calibration is correct.
Содержание Flo-Gard 6301
Страница 8: ...vi FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 12: ...x FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 20: ...1 General Information 1 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 30: ...2 Controls and Indicators 2 10 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 34: ...3 Administration Sets 3 4 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 36: ...4 Nurse Call 4 2 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 74: ...7 Troubleshooting Chart 7 28 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 110: ...8 Disassembly Reassembly 8 36 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 200: ...13 Overview 13 2 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 201: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 3 Figure 13 1 System Block Diagram ...
Страница 205: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 7 Figure 13 5 Power Supply Section ...
Страница 206: ...13 8 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 6 DC Power Supply ...
Страница 210: ...13 12 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 10 Master CPU Schematic ...
Страница 213: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 15 Figure 13 13 Address Decoder ...
Страница 216: ...13 18 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 16 Slave CPU Schematic ...
Страница 220: ...13 22 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 20 CPU PCB Sensor Block ...
Страница 221: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 23 Figure 13 21 PPI Block 1 ...
Страница 222: ...13 24 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 22 PPI Block 2 ...
Страница 224: ...13 26 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 24 Air Sensing Block ...
Страница 226: ...13 28 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 26 Display PCB ...
Страница 229: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 31 Figure 13 29 LCDs for Pump 1 ...
Страница 230: ...13 32 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 30 LCDs for Pump 2 ...
Страница 231: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 33 Figure 13 31 LED Lamps ...
Страница 232: ...13 34 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 Figure 13 32 Audible Alarm PCB ...
Страница 233: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 35 Figure 13 33 Backup Buzzer PCB ...
Страница 234: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 36 Figure 13 34 Terminal PCB ...
Страница 235: ...07 19 B1 729 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 13 37 Figure 13 35 Accessories ...
Страница 244: ...A A 6 FLO GARD 6301 DUAL CHANNEL VOLUMETRIC INFUSION PUMP SERVICE MANUAL 07 19 B1 729 ...
Страница 251: ...07 19 B1 729 ...