4.2 Back Panel
The following photograph illustrates the back of the Site~Rite IV Scanner.
The following table lists and describes the back panel features:
Site~Rite* IV Ultrasound System Operator’s Manual
Chapter 4: Site~Rite* IV Scanner
Controls
Gain Control
Buttons
6
Description
Controls the gain, or intensity, of the scan image. Reduce gain
by pressing on the left-hand side of button. Increase gain by
pressing the right-hand side of button.
The gain level is displayed as a marker on a horizontal bar at the
bottom of the screen, with the least amount of gain on the left-
hand side, and greatest amount of gain on the right-hand side.
By default, the gain is set to the middle setting.
Increase Brightness
Button
7
Controls the brightness of the display. Increase brightness by
pressing on the button.
2
3
1
4
5
6
8
Decrease Brightness
Button
Controls the brightness of the display. Decrease brightness by
pressing on the button.
Feature
Fuse
1
Description
Location of External 5 A 125 V Micro
TM
Fuse.
Power Source
Connector
2
Use the Power Source Connector to connect a battery pack or
an A/C adapter to the Site~Rite IV Scanner. See Chapter 3.4,
Attaching Power Sources.
Battery
Brackets
3
Brackets to which the battery is attached by its clips.
4
5
6
Probe Port
Equipotential
Connector
Video Out
Use the probe port to connect a Site~Rite IV Probe
to the scanner.
Equipotential Ground Connection
Use the Video Out Port to attach the Site~Rite IV Scanner to
external video equipment or printers.