Site~Rite* IV Ultrasound System Operator’s Manual

7.2  Charging the Battery with the Site~Rite

*

 Mini Battery 

Charger

Site~Rite IV Battery Packs are fully charged prior to shipment, but may require charging before use. 

To charge a battery pack with the Site~Rite Mini Battery Charger,

1. 

Plug the Site~Rite Mini Battery Charger into a power outlet.

2. 

Verify that the LED indicator on the front of the Site~Rite Mini Battery Charger is 
not illuminated.

3. 

Attach the battery pack to the Site~Rite Mini Battery Charger. The battery packs attach to the 
Site~Rite  Mini  Battery  Charger  by  inserting  the  molded  battery  charger  connector  into  the 
matching connector of the battery pack.
• 

Hold the battery pack with the plug port visible.

• 

Place the molded battery connector of the Site~Rite Mini Battery Charger into 
the plug port.

• 

The molded battery connector will lock into the battery pack port.

• 

Gently  tug  on  the  pack  to  verify  that  it  is  securely  connected  to  the  Site~Rite 
Mini Battery Charger.

4. 

The LED indicator will flash green to indicate that the charger is in Rapid Charge. 

 

After  approximately  3  hours,  the  LED  indicator  will  glow  solid  green  to  indicate 
charge complete.

 

Following Rapid Charge the battery charger will continue to provide a trickle/top-off charge for 
4 hours. The battery charger will then provide maintenance charge to the battery for 
a 10 hour period.  

Note: 

When placing a cold battery in a warm room to charge, the rapid temperature change of the 
battery occasionally causes the Site~Rite Mini Battery Charger to switch from the Rapid Charge 
mode to Pending Charge mode (within a few minutes of turning on). If this occurs, remove the 
Site~Rite Mini Battery Charger from the AC power and remove the battery from the molder bat-
tery charger connector; and wait a few moments before reconnecting the Site~Rite Mini Battery 
Charger to AC power and the battery to the molded battery charger connector.

To remove a battery pack from Site~Rite Mini Battery Charger,

1. 

Press down on the molder battery charger connector spring lock, and gently pull the battery 
pack off of the connector.

You  can  remove  the  Site~Rite  Mini  Battery  Charger  from  AC  power  and  remove  the  battery  for  use 
when the Rapid Charge indicator glows solid green. At the end of the Rapid Charge cycle, the battery is 
approximately 85% charged. The Site~Rite Mini Battery Charger then switches to the Pending Charge 
mode to continue the charging process while preventing the battery from overheating. Using a partially 
charged battery does not create battery memory or reduce the life of the battery.

No harm occurs if you leave the battery on the Site~Rite Mini Battery Charger after it is fully charged. 

Chapter 7: Site~Rite* IV Battery Charger

Содержание Site-Rite IV

Страница 1: ...ent two years have elapsed between this date and product use the user should contact Bard Access Systems Inc to see if additional product information is available Revision date January 2007 Bard Charg...

Страница 2: ...Site Rite IV Ultrasound System Operator s Manual...

Страница 3: ......

Страница 4: ...eeze Image Do not dispose with ordinary municipal waste European Conformity Unfreeze Image Medical Electrical Equipment Classified by UL with respect to Electrical Shock Fire and Mechanical Hazards on...

Страница 5: ...ite IV Probes 5 4 Cleaning and Disinfecting the Scanner and Probes 5 5 Periodic Probe Testing 6 Site Rite IV Needle Guides 6 1 Attaching the Needle Guide to the Probe 6 2 Using the Needle Guides 7 Sit...

Страница 6: ...any modification to the product except those made by the manufacturer is strictly prohibited by law Federal U S A law restricts this device to sale by or on the order of physician This product should...

Страница 7: ...ite Mini Battery Charger Connecting the Site Rite IV Scanner to any A C adapter other than the one provided with the scanner Cautions Caution When using Site Rite IV Probes in a sterile environment wi...

Страница 8: ...pproved for use by country and local authorities are to be connected to the Site Rite IV Scanner Use of accessories that are not compliant to International Electrical Safety Standards may result in de...

Страница 9: ...victim is conscious immediately give 2 to 4 glasses of water and induce vomiting by touching finger to back of throat Get medical attention immediately Supplied To Bard Access Systems Inc Salt Lake C...

Страница 10: ...n all of the screws There are two screws per mounting bracket 3 Attaching the Side Arm To attach the side arm 1 Use the Allen wrench to unscrew the clamp screws 2 Position and hold the side arm on the...

Страница 11: ...of the mounting plate 7 The Site Rite IV Scanner is now attached to the roll stand If the scanner is temporarily removed from the roll stand use the scanner by placing it on a horizontal flat surface...

Страница 12: ...he battery pack see Chapter 7 Site Rite Battery Charger Note If you are using the A C adapter the adapter automatically adjusts to the correct line voltage between 100 240 VAC 50 60 Hz The adapter shi...

Страница 13: ...be connector 3 Gently pull the connector away from the back of Site Rite IV Scanner For information about probe types and handling see Chapter 5 Site Rite IV Probes For more information about probe sp...

Страница 14: ...depths see Chapter 5 Site Rite IV Probes Displays or removes dot markers that overlay the center of the image Dot markers indicate scan depth from the surface of the skin By default dot markers are n...

Страница 15: ...default the gain is set to the middle setting Increase Brightness Button 7 Controls the brightness of the display Increase brightness by pressing on the button 2 3 1 4 5 6 8 Decrease Brightness Butto...

Страница 16: ...the left hand side of the screen The default date format is mm dd yy 8 To save the changes simultaneously press the Depth Selection and Image Reversal buttons 9 Press the Freeze Frame button to return...

Страница 17: ...of 4 cm The label on the connector of the probe indicates the probe s frequency the focal length and the probe s serial number Each probe has two scan depths that you can select using the Site Rite IV...

Страница 18: ...ientation Standoff Probe red and blue cable bend relief When using a 7 5 MHz probe red cable bend relief or 9 0 MHz probe blue cable bend relief hold the probe so that the side with the needle guide h...

Страница 19: ...th over the areas of the roll stand requiring cleaning Note It is important to remove all particles or other matter from all surfaces and crevices Recommendations for Cleaning the Site Rite IV Probes...

Страница 20: ...probe for proper alignment 1 Connect the probe to the Site Rite IV Scanner 2 Press the Power button 3 Press the Dot Marker button to display the center dot markers on the screen 4 Clip 1 5 cm Site Rit...

Страница 21: ...markers the probe is aligned properly If the needle displays to either side of the dot markers the probe is misaligned and must be returned to Bard Access Systems for repair For information about ret...

Страница 22: ...Guide to the Probe To attach the needle guide to a probe 1 Choose the appropriate needle guide based on the depth of the structure to be punctured and the gauge of the needle to be used Note Needle gu...

Страница 23: ...larger end of the needle guide toward the probe until the needle guide snaps on to the needle guide hook Do not slide Caution Always snap the needle guides on to the Probe hook Do not slide the needle...

Страница 24: ...s centered on the target vessel slowly advance the needle while looking at the screen of the Site Rite IV scanner When the needle approaches the target vessel you should see the anterior wall indentin...

Страница 25: ...ery Charger does not have an external fuse Warning Use only the Site Rite Mini Battery Charger or Charge Rite Battery Charger to charge Site Rite IV Battery Packs Use of any other charger to charge Si...

Страница 26: ...arge for 4 hours The battery charger will then provide maintenance charge to the battery for a 10 hour period Note When placing a cold battery in a warm room to charge the rapid temperature change of...

Страница 27: ...your warranty Never connect a Site Rite IV A C Adapter to a Charge Rite Battery Charger or Site Rite Mini Battery Charger 7 4 Indicator Lights The following table lists and describes the indicator li...

Страница 28: ...it a few moments before turning Charge Rite Battery Charger back on To remove a battery pack from Charge Rite Battery Charger 1 On the back of Charge Rite Battery Charger turn the power switch to Off...

Страница 29: ...rly connected to scanner Probe not properly connected or not connected at all Probe damaged Remove battery pack and replace with a fully charged battery pack Verify that power source is firmly connect...

Страница 30: ...operly connect the Probe Assembly The Probe Assembly is defective Replace the Probe Assembly with a known functioning assembly Return the probe for repair Scanner turns on screen lights up text is dis...

Страница 31: ...Access Systems in US or local distribu tor outside of US in US or local distributor outside of US Invalid Probe No Probe Invalid Depth Turn scanner off Allow system to return to room temperature and t...

Страница 32: ...one provided with the scanner This Limited product warranty is in lieu of all other warranties express or implied including without limitation any warranty of merchantability or fitness for a particul...

Страница 33: ...10 8 cm H x 1 75 4 5 cm L Weight 1 9 lbs 0 86 kg Battery Type Nickel Metal Hydride Output Voltage 14 4 VDC Duration of Site Rite IV Operation when fully charged 2 hours AC Adapter Specifications Dime...

Страница 34: ...be performed using the principle of ALARA As Low As Reasonably Achievable The Site Rite IV System holds the acoustic output of the device constant The licensed medical practitioner should limit the t...

Страница 35: ...ministration in the United States of America as well as other agencies throughout the world This product is ISO certified and CE approved Therefore any modification to the product except those made by...

Страница 36: ...90 Clinical Information 800 443 3385 Fax 801 595 4948 www bardaccess com Authorised European Service Representative Bard Limited Forest House Brighton Road Crawley West Sussex RH11 9BP England Phone 4...

Отзывы: