Site~Rite* IV Ultrasound System Operator’s Manual
Chapter 5: Site~Rite* IV Probes
Site~Rite
*
IVProbes
5.1 Overview
Standoff Probe
The following table describes the probe.
For superficial imaging, there are two standoff probes:
•
7.5 MHz probe, with a red cable bend relief, optimized for imaging at depths from 1.5 cm to
4 cm below skin surface. The probe images to a depth of 4 cm.
•
9.0 MHz probe, with a blue cable bend relief, optimized for imaging at depths from 0 cm to
1.5 cm below skin surface. The probe images to a depth of 4 cm.
The label on the connector of the probe indicates the probe’s frequency, the focal length, and the
probe’s serial number.
Each probe has two scan depths that you can select using the Site~Rite IV Scanner Depth Selection
button.
The following table lists the probe type and bend relief color, scan depths, and scan plane shape:
*Default settings are bold.
For more information about Site~Rite IV Scanner functions, see Chapter 4,
Site~Rite IV Scanner
.
2
1
Feature
Standoff
Probe
1
Description
Needle Guide
Hook
2
Used for superficial imaging to a depth of 4 cm.
Hook to which the short end of the needle guide is clipped.
7.5 MHz
Standoff
(Red)
2 cm
9.0 MHz
Standoff
(Blue)
4 cm
2 cm
4 cm
Trapezoid
Trapezoid
Scan Depth
Setting*
Image Shape