8.1 Operating and Storage Conditions
The Site~Rite IV System should not be operated or stored in conditions beyond the following tolerances:
8.2 Troubleshooting
If you have problems with the operation of your Site~Rite IV System, review this section before
contacting Bard Access Systems.
Common Problems and Solutions
The following table describes common problems, possible causes, and suggested solutions:
Site~Rite* IV Ultrasound System Operator’s Manual
Chapter 8: Site~Rite* IV System Maintenance and Troubleshooting
Problem
Possible Cause
Scanner does not turn on.
Battery pack is depleted.
Battery pack or A/C adaptor
not properly connected to
scanner.
Probe not properly connected
or not connected at all.
Probe damaged.
Remove battery pack and
replace with a fully charged
battery pack.
Verify that power source is
firmly connected to back of
scanner.
Verify that probe is properly
connected to scanner.
Attach different Site~Rite IV
Probe and power on. If
scanner functions, return
probe for repair. If scanner
does not function, return
scanner for repair.
Solution
A/C adaptor is not properly
attached to an outlet.
Verify that the A/C adaptor
is properly attached to a
functional outlet.
The Battery Pack is defective.
Replace the Battery Pack.
Condition
59°F to 100°F
(15°C to 38°C)
Tolerance
41°F to 113°F
(5°C to 45°C)
5% to 95% Relative Humidity (non-condensing)
-500 feet to 10,000 feet elevation
(103 kPA to 70 kPA)
Operating Temperature:
Storage Temperature:
Operating and Storage Humidity:
Operating and Storage Atmospheric
Pressure: