Site~Rite* IV Ultrasound System Operator’s Manual
5.4 Cleaning and Disinfecting the Scanner and Probes
Warning:
Do not attempt to sterilize the Site~Rite IV Scanner or Probes with ethylene oxide or heat
sterilization methods. Damage to the devices will result.
Recommendations for Cleaning the Site~Rite IV Scanner
To clean the scanner:
1.
Dampen a nonabrasive cloth with either warm water or rubbing alcohol.
2.
Gently wipe the dampened cloth over the areas of the scanner requiring cleaning.
Note:
When cleaning the system and components, it is important to remove all particles or
other matter from all surfaces and crevices.
Warning:
Do not let liquid leak into the scanner, Charge~Rite Battery Charger, Site~Rite Mini
Battery Charger, Battery Packs, A/C Adapter, Probe Connector, or Probe Port.
Recommendations for Cleaning the Charge~Rite Battery Charger and
Site~Rite Mini Battery Charger
To clean the battery charger:
1.
Dampen a nonabrasive cloth with either warm water or rubbing alcohol.
2.
Gently wipe the dampened cloth over the areas of the battery charger requiring cleaning.
Note:
It is important to remove all particles or other matter from all surfaces and crevices.
Warning:
Do not let liquid leak into the scanner, Charge~Rite Battery Charger, Site~Rite Mini
Battery Charger, Battery Packs, A/C Adapter, Probe Connector, or Probe Port.
Recommendations for Cleaning the Site~Rite Roll Stand
To clean the roll stand:
1.
Dampen a nonabrasive cloth with either warm water or rubbing alcohol.
2.
Gently wipe the dampened cloth over the areas of the roll stand requiring cleaning.
Note:
It is important to remove all particles or other matter from all surfaces and crevices.
Recommendations for Cleaning the Site~Rite IV Probes
The probes should be cleaned and disinfected after each use.
To clean the probes:
1.
Dampen a nonabrasive cloth with either warm water or rubbing alcohol.
2.
Gently wipe the dampened cloth over the probe head and handle.
Note:
It is important to remove all particles of adhering tissue, ultrasound gel, dried blood, or
other matter from all surfaces and crevices.
Note:
Immediately after cleaning, the probes should be rinsed thoroughly with warm water or
rubbing alcohol.
Caution:
Use only warm water or rubbing alcohol for Probe cleaning. Use of hot water
(temperature exceeding 113°F/45°C) may damage the probe.
Warning:
Do not let liquid leak into the scanner, Charge~Rite Battery Charger, Site~Rite Mini
Battery Charger, Battery Packs, A/C Adapter, Probe Connector, or Probe Port.
Recommendations for Liquid Disinfection of the Site~Rite IV Probes
The Site~Rite IV Probes may be liquid disinfected by soaking them in Cidex
*
plus 28 day solution
available from Johnson and Johnson Medical, Inc.
Follow the solution manufacturer’s recommendations for soak time necessary to achieve the desired
germicide level of activity.
Caution:
During disinfection immerse the probe up to, but not including, the colored
cable bend relief.
Warning:
Do not soak the probe cable, cable bend reliefs, or probe connectors in Cidex solution.
Damage to the probe will result.
Warning:
Do not let liquid leak into the scanner, Charge~Rite Battery Charger, Site~Rite Mini
Battery Charger, Battery Packs, A/C Adapter, Probe Connector, or Probe Port.