Site~Rite* IV Ultrasound System Operator’s Manual

Chapter 1: Overview

Warning: 

When using Site~Rite Needle Guides on the fluid standoff probes, use only sterile  
plastic sheaths that are 1 mil (0.001 inch or 0.0254 mm) thick. 

Warning: 

Do not let liquid leak into the scanner, Charge~Rite Battery Charger, Site~Rite Mini 
Battery Charger, Battery Packs, A/C Adapter, Probe Connector or Probe Port.

Warning: 

Do not attempt to sterilize the Site~Rite IV Scanner or Probes with ethylene oxide or heat 
sterilization methods. Damage to the devices will result.

Warning: 

Do not soak the probe cable, cable bend reliefs or probe connectors in Cidex solution. 
Damage to the probe will result.

Warning: 

Never use probes that are out of alignment for needle guidance. 

Warning: 

Always properly dispose of dead battery packs. Never incinerate a battery pack. If necessary, 
return Site~Rite IV Battery Packs to Bard Access Systems, for disposal.

Warning: 

Only qualified personnel should attempt to service this equipment.

Warning: 

The Site~Rite IV System contains dangerous voltages. Service attempts by unqualified  
personnel may result in serious injury or death.

Warning: 

The Site~Rite IV contains static sensitive components and circuits. Failure to observe  
proper static control procedures may result in damage to the system.

Warning: 

The following actions void the warranty of the Site~Rite IV System.

• 

Opening or servicing the scanner or the probe housing.

• 

Removal of system labels by anyone other than by Bard Access Systems, authorized service 
personnel.

• 

Opening or servicing the battery pack, A/C adapter, Charge~Rite Battery Charger, or the 
Site~Rite Mini Battery Charger by anyone other than Bard Access Systems authorized ser-
vice personnel.

• 

Connecting the Site~Rite IV Scanner to any battery source other than the Site~Rite IV 
Battery Pack.

• 

Connecting the Site~Rite IV Battery Packs to any charger other than the Charge~Rite 
Battery Charger or the Site~Rite Mini Battery Charger.

• 

Connecting the Site~Rite IV Scanner to any A/C adapter other than the one provided 
with the scanner. 

Cautions

Caution:  

When using Site~Rite IV Probes in a sterile environment with a Site~Rite Disposable Needle 
Guide, use only sterile plastic sheaths that are 1 mil (0.001 inch or 0.0254 mm) thick.

Caution:  

Always inspect the ultrasound probe cap, cover and cable before using it with the 
Site~Rite IV Scanner. Do not use the probe if the cap is cracked or damaged. Do not 
allow the probe liquid or liquid vapor to contact the eyes, skin, clothing or room furnish-
ings. The fluid can be irritating to the skin and eyes and may damage furniture finishes. In 
case of contact, immediately rinse the area with water.

Caution:  

Do not scratch the acoustic window (or face) of the probe. Image quality may be  
degraded if the probe acoustic window (or face) is scratched.

Caution:  

Do not pull the cable to disconnect the probe-connector from the scanner. Pulling the 
cable may destroy circuit continuity at the connector junction, damage the cable or  
cable-connector or possibly damage circuit continuity within the scanner.

Caution:  

Do not twist or bend the cable in excess of that required during normal use of the probe. 
Excessive twisting or bending of the cable may cause failure or intermittent or unpredict-
able operation. 

Caution:  

Do not disinfect the probes using gas or heat methods. Never autoclave a probe.

Caution:  

When disinfecting the probes with Cidex

*

, do not soak the probe cable, cable bend relief 

or probe connector in Cidex

*

. Doing so damages the probe.

Caution:  

Hot water (in excess of 113°F or 45°C) may damage the probe.

Caution:  

Do not expose the probe to direct sunlight for an extended period of time so it does not 
absorb excessive heat from the sun and damage the probe.

Caution:  

Do not let liquid leak into the Site~Rite IV Scanner, Charge~Rite Battery Charger, Site~Rite 
Mini Battery Charger, Battery Packs, A/C Adapter, Probe Connector or Probe Port.

Caution:  

Before you connect the A/C adapter to or disconnect the A/C adapter from the 
Site~Rite IV Scanner, verify that the A/C adapter is not plugged into a wall socket.

­

Содержание Site-Rite IV

Страница 1: ...ent two years have elapsed between this date and product use the user should contact Bard Access Systems Inc to see if additional product information is available Revision date January 2007 Bard Charg...

Страница 2: ...Site Rite IV Ultrasound System Operator s Manual...

Страница 3: ......

Страница 4: ...eeze Image Do not dispose with ordinary municipal waste European Conformity Unfreeze Image Medical Electrical Equipment Classified by UL with respect to Electrical Shock Fire and Mechanical Hazards on...

Страница 5: ...ite IV Probes 5 4 Cleaning and Disinfecting the Scanner and Probes 5 5 Periodic Probe Testing 6 Site Rite IV Needle Guides 6 1 Attaching the Needle Guide to the Probe 6 2 Using the Needle Guides 7 Sit...

Страница 6: ...any modification to the product except those made by the manufacturer is strictly prohibited by law Federal U S A law restricts this device to sale by or on the order of physician This product should...

Страница 7: ...ite Mini Battery Charger Connecting the Site Rite IV Scanner to any A C adapter other than the one provided with the scanner Cautions Caution When using Site Rite IV Probes in a sterile environment wi...

Страница 8: ...pproved for use by country and local authorities are to be connected to the Site Rite IV Scanner Use of accessories that are not compliant to International Electrical Safety Standards may result in de...

Страница 9: ...victim is conscious immediately give 2 to 4 glasses of water and induce vomiting by touching finger to back of throat Get medical attention immediately Supplied To Bard Access Systems Inc Salt Lake C...

Страница 10: ...n all of the screws There are two screws per mounting bracket 3 Attaching the Side Arm To attach the side arm 1 Use the Allen wrench to unscrew the clamp screws 2 Position and hold the side arm on the...

Страница 11: ...of the mounting plate 7 The Site Rite IV Scanner is now attached to the roll stand If the scanner is temporarily removed from the roll stand use the scanner by placing it on a horizontal flat surface...

Страница 12: ...he battery pack see Chapter 7 Site Rite Battery Charger Note If you are using the A C adapter the adapter automatically adjusts to the correct line voltage between 100 240 VAC 50 60 Hz The adapter shi...

Страница 13: ...be connector 3 Gently pull the connector away from the back of Site Rite IV Scanner For information about probe types and handling see Chapter 5 Site Rite IV Probes For more information about probe sp...

Страница 14: ...depths see Chapter 5 Site Rite IV Probes Displays or removes dot markers that overlay the center of the image Dot markers indicate scan depth from the surface of the skin By default dot markers are n...

Страница 15: ...default the gain is set to the middle setting Increase Brightness Button 7 Controls the brightness of the display Increase brightness by pressing on the button 2 3 1 4 5 6 8 Decrease Brightness Butto...

Страница 16: ...the left hand side of the screen The default date format is mm dd yy 8 To save the changes simultaneously press the Depth Selection and Image Reversal buttons 9 Press the Freeze Frame button to return...

Страница 17: ...of 4 cm The label on the connector of the probe indicates the probe s frequency the focal length and the probe s serial number Each probe has two scan depths that you can select using the Site Rite IV...

Страница 18: ...ientation Standoff Probe red and blue cable bend relief When using a 7 5 MHz probe red cable bend relief or 9 0 MHz probe blue cable bend relief hold the probe so that the side with the needle guide h...

Страница 19: ...th over the areas of the roll stand requiring cleaning Note It is important to remove all particles or other matter from all surfaces and crevices Recommendations for Cleaning the Site Rite IV Probes...

Страница 20: ...probe for proper alignment 1 Connect the probe to the Site Rite IV Scanner 2 Press the Power button 3 Press the Dot Marker button to display the center dot markers on the screen 4 Clip 1 5 cm Site Rit...

Страница 21: ...markers the probe is aligned properly If the needle displays to either side of the dot markers the probe is misaligned and must be returned to Bard Access Systems for repair For information about ret...

Страница 22: ...Guide to the Probe To attach the needle guide to a probe 1 Choose the appropriate needle guide based on the depth of the structure to be punctured and the gauge of the needle to be used Note Needle gu...

Страница 23: ...larger end of the needle guide toward the probe until the needle guide snaps on to the needle guide hook Do not slide Caution Always snap the needle guides on to the Probe hook Do not slide the needle...

Страница 24: ...s centered on the target vessel slowly advance the needle while looking at the screen of the Site Rite IV scanner When the needle approaches the target vessel you should see the anterior wall indentin...

Страница 25: ...ery Charger does not have an external fuse Warning Use only the Site Rite Mini Battery Charger or Charge Rite Battery Charger to charge Site Rite IV Battery Packs Use of any other charger to charge Si...

Страница 26: ...arge for 4 hours The battery charger will then provide maintenance charge to the battery for a 10 hour period Note When placing a cold battery in a warm room to charge the rapid temperature change of...

Страница 27: ...your warranty Never connect a Site Rite IV A C Adapter to a Charge Rite Battery Charger or Site Rite Mini Battery Charger 7 4 Indicator Lights The following table lists and describes the indicator li...

Страница 28: ...it a few moments before turning Charge Rite Battery Charger back on To remove a battery pack from Charge Rite Battery Charger 1 On the back of Charge Rite Battery Charger turn the power switch to Off...

Страница 29: ...rly connected to scanner Probe not properly connected or not connected at all Probe damaged Remove battery pack and replace with a fully charged battery pack Verify that power source is firmly connect...

Страница 30: ...operly connect the Probe Assembly The Probe Assembly is defective Replace the Probe Assembly with a known functioning assembly Return the probe for repair Scanner turns on screen lights up text is dis...

Страница 31: ...Access Systems in US or local distribu tor outside of US in US or local distributor outside of US Invalid Probe No Probe Invalid Depth Turn scanner off Allow system to return to room temperature and t...

Страница 32: ...one provided with the scanner This Limited product warranty is in lieu of all other warranties express or implied including without limitation any warranty of merchantability or fitness for a particul...

Страница 33: ...10 8 cm H x 1 75 4 5 cm L Weight 1 9 lbs 0 86 kg Battery Type Nickel Metal Hydride Output Voltage 14 4 VDC Duration of Site Rite IV Operation when fully charged 2 hours AC Adapter Specifications Dime...

Страница 34: ...be performed using the principle of ALARA As Low As Reasonably Achievable The Site Rite IV System holds the acoustic output of the device constant The licensed medical practitioner should limit the t...

Страница 35: ...ministration in the United States of America as well as other agencies throughout the world This product is ISO certified and CE approved Therefore any modification to the product except those made by...

Страница 36: ...90 Clinical Information 800 443 3385 Fax 801 595 4948 www bardaccess com Authorised European Service Representative Bard Limited Forest House Brighton Road Crawley West Sussex RH11 9BP England Phone 4...

Отзывы: