Site~Rite* IV Ultrasound System Operator’s Manual
Chapter 1: Overview
Warning:
When using Site~Rite Needle Guides on the fluid standoff probes, use only sterile
plastic sheaths that are 1 mil (0.001 inch or 0.0254 mm) thick.
Warning:
Do not let liquid leak into the scanner, Charge~Rite Battery Charger, Site~Rite Mini
Battery Charger, Battery Packs, A/C Adapter, Probe Connector or Probe Port.
Warning:
Do not attempt to sterilize the Site~Rite IV Scanner or Probes with ethylene oxide or heat
sterilization methods. Damage to the devices will result.
Warning:
Do not soak the probe cable, cable bend reliefs or probe connectors in Cidex solution.
Damage to the probe will result.
Warning:
Never use probes that are out of alignment for needle guidance.
Warning:
Always properly dispose of dead battery packs. Never incinerate a battery pack. If necessary,
return Site~Rite IV Battery Packs to Bard Access Systems, for disposal.
Warning:
Only qualified personnel should attempt to service this equipment.
Warning:
The Site~Rite IV System contains dangerous voltages. Service attempts by unqualified
personnel may result in serious injury or death.
Warning:
The Site~Rite IV contains static sensitive components and circuits. Failure to observe
proper static control procedures may result in damage to the system.
Warning:
The following actions void the warranty of the Site~Rite IV System.
•
Opening or servicing the scanner or the probe housing.
•
Removal of system labels by anyone other than by Bard Access Systems, authorized service
personnel.
•
Opening or servicing the battery pack, A/C adapter, Charge~Rite Battery Charger, or the
Site~Rite Mini Battery Charger by anyone other than Bard Access Systems authorized ser-
vice personnel.
•
Connecting the Site~Rite IV Scanner to any battery source other than the Site~Rite IV
Battery Pack.
•
Connecting the Site~Rite IV Battery Packs to any charger other than the Charge~Rite
Battery Charger or the Site~Rite Mini Battery Charger.
•
Connecting the Site~Rite IV Scanner to any A/C adapter other than the one provided
with the scanner.
Cautions
Caution:
When using Site~Rite IV Probes in a sterile environment with a Site~Rite Disposable Needle
Guide, use only sterile plastic sheaths that are 1 mil (0.001 inch or 0.0254 mm) thick.
Caution:
Always inspect the ultrasound probe cap, cover and cable before using it with the
Site~Rite IV Scanner. Do not use the probe if the cap is cracked or damaged. Do not
allow the probe liquid or liquid vapor to contact the eyes, skin, clothing or room furnish-
ings. The fluid can be irritating to the skin and eyes and may damage furniture finishes. In
case of contact, immediately rinse the area with water.
Caution:
Do not scratch the acoustic window (or face) of the probe. Image quality may be
degraded if the probe acoustic window (or face) is scratched.
Caution:
Do not pull the cable to disconnect the probe-connector from the scanner. Pulling the
cable may destroy circuit continuity at the connector junction, damage the cable or
cable-connector or possibly damage circuit continuity within the scanner.
Caution:
Do not twist or bend the cable in excess of that required during normal use of the probe.
Excessive twisting or bending of the cable may cause failure or intermittent or unpredict-
able operation.
Caution:
Do not disinfect the probes using gas or heat methods. Never autoclave a probe.
Caution:
When disinfecting the probes with Cidex
*
, do not soak the probe cable, cable bend relief
or probe connector in Cidex
*
. Doing so damages the probe.
Caution:
Hot water (in excess of 113°F or 45°C) may damage the probe.
Caution:
Do not expose the probe to direct sunlight for an extended period of time so it does not
absorb excessive heat from the sun and damage the probe.
Caution:
Do not let liquid leak into the Site~Rite IV Scanner, Charge~Rite Battery Charger, Site~Rite
Mini Battery Charger, Battery Packs, A/C Adapter, Probe Connector or Probe Port.
Caution:
Before you connect the A/C adapter to or disconnect the A/C adapter from the
Site~Rite IV Scanner, verify that the A/C adapter is not plugged into a wall socket.