Site~Rite* IV Ultrasound System Operator’s Manual
Chapter 3: Setting up the Site~Rite* IV System
3.4 Attaching Feet to the Site~Rite
*
IV Scanner (Optional)
The Site~Rite IV Scanner comes with four black rubber feet that can be attached to the bottom of the
scanner to allow for tabletop use. The four black rubber feet are included in an individual package with
the system.
Since all Site~Rite IV Ultrasound Systems come with a roll-stand the rubber feet are not installed.
The rubber feet can be easily installed on the scanner. Remove the screw in each corner of the scanner.
Insert the screw into one of the black rubber feet. Replace the foot and screw back into the corner of
the scanner, tightening the screw until secure.
3.5 Attaching Power Sources
The Site~Rite IV System can be powered by either a nickel-metal hydride battery pack or an A/C
adapter. Both the battery pack and the A/C adapter attach in the same manner.
For more information about the battery pack, see Chapter 7,
Site~Rite Battery Charger
.
Note:
If you are using the A/C adapter, the adapter automatically adjusts to the correct
line voltage between 100 - 240 VAC, 50 - 60 Hz. The adapter ships with country-specific
wall plugs. Verify that the A/C adapter has the appropriate plug for your country’s
electrical system.
To connect a power source (battery pack or A/C adapter) to the Site~Rite IV Scanner,
1.
Hold the power source so that the power connector projection points down.
2.
Align the power source with the scanner’s power source connector.
3.
Press together the spring clips located on the top and bottom of the power source.
4.
Place the power source against the back of the Site~Rite IV Scanner, between the upper and
lower guides, and release the clips. The clips snap into the notches on the guides.
5.
Gently tug on the power source to verify that it is securely connected to Site~Rite IV Scanner.
6.
If you connected an A/C adapter to your Site~Rite IV Scanner, plug the power cord into a
wall socket.
To remove a power source from the Site~Rite IV Scanner, press down on the spring clips, and gently
pull the power source off of the scanner.