Chapter 8: DoseCom™
951128 Rev E (11/10)
75
sequential delivery. Changes in rate or dose could be an indication of a titration. Titration
checking provides a means to confirm the proper rate or dose for the titrated medication. The
process is completed as follows:
1. Transmit the infusion data to the pump as described in the previous sections.
2. Enter a valid, non-zero value for rate or dose if applicable.
3. Validate the transmitted infusion data by moving the Cursor to Accept and press
.
4. The titration checking routine commences as follows:
For Standard Mode Deliveries
• If the Patient ID and Drug Name match the previous infusion information, the rate
value is compared.
• If the rate values are equal, the rate and
VTBD are displayed as transmitted and the
infusion can be started by pressing
.
• If the rate values are not equal, the following
Titration Alert will be displayed:
1. Use the
to highlight the cor-
rect rate, then press
.
2. Begin the infusion by pressing
.
For Dose Mode Deliveries
• If the Patient ID, Drug Name, Dose units, and Drug Concentration match the previous
infusion information, the dose value is compared.
• If the dose values are equal, the infusion
information is displayed and the infusion can
be started by pressing
.
• If the dose values are not equal, the following
Titration Alert will be displayed:
1. Use the
to highlight the cor-
rect dose, then press
.
2. Begin the infusion by pressing
.
Содержание OUTLOOK ES Series
Страница 2: ...951128 Rev E 11 10 ...
Страница 4: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 30: ...OUTLOOK ES OPERATOR S MANUAL 22 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 90: ...OUTLOOK ES OPERATOR S MANUAL 82 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 96: ...OUTLOOK ES OPERATOR S MANUAL 88 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 102: ...OUTLOOK ES OPERATOR S MANUAL 94 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 131: ...951128 Rev E 11 10 longer intervals of time Trumpet Curve 2nd Hour of Delivery Trumpet Curve Last Hour of Delivery ...
Страница 132: ...OUTLOOK ES OPERATOR S MANUAL 124 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 133: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...