About the Pump
951128 Rev E (11/10)
1
A
BOUT
THE
P
UMP
INTENDED USE
The Outlook® ES is intended for use with B. Braun Medical Inc. Horizon® Pump IV Sets to
regulate the flow of primary and secondary fluids when positive pressure is required. The
infusion system is capable of delivering fluid from a negative head height (when the IV fluid
container is lower than the pump), and provides clinically accepted volumetric accuracy for
all standard IV fluids, including blood, lipids, and Total Parenteral Nutrition (TPN).
Positive pressure is frequently a necessity, but clinical experience shows that high pressure
limits may increase the severity of an infiltration without causing an alarm. Because there is a
need to control pump pressure settings, the Outlook ES has user-selectable Occlusion Limit
settings which start at 75 mmHg and extend to 500 mmHg. At rates of 400 ml/hr and higher,
the pressure setting is automatically increased to 400 mmHg. For epidural infusions which
require higher pressures, the pump has an extended occlusion limit setting of 750 mmHg.
The pump is equipped with distinct audible and visual alarm signals to indicate Keep Vein
Open (KVO), low battery, and other alarm conditions.
EPIDURAL ADMINISTRATION
The Outlook ES can be used for epidural administration of anesthetic and analgesic drugs.
For epidural administration of anesthetic drugs, use indwelling catheters specifically indicated
for short-term (96 hours or less) anesthetic epidural drug delivery.
For epidural administration of analgesic drugs, use indwelling catheters specifically indicated
for either short-term or long-term analgesic epidural drug delivery.
When the Outlook ES is being used for epidural infusions, the source container and
administration set should be clearly differentiated from source containers and administration
sets used for other routes of infusion.
Following an occlusion alarm, a small amount of fluid may be delivered. Use of a microbore
set assures this is less than 0.1 ml.
The administration of drugs is restricted to those anesthetic and analgesic drugs labeled for
continuous epidural administration.
Содержание OUTLOOK ES Series
Страница 2: ...951128 Rev E 11 10 ...
Страница 4: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 30: ...OUTLOOK ES OPERATOR S MANUAL 22 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 90: ...OUTLOOK ES OPERATOR S MANUAL 82 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 96: ...OUTLOOK ES OPERATOR S MANUAL 88 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 102: ...OUTLOOK ES OPERATOR S MANUAL 94 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 131: ...951128 Rev E 11 10 longer intervals of time Trumpet Curve 2nd Hour of Delivery Trumpet Curve Last Hour of Delivery ...
Страница 132: ...OUTLOOK ES OPERATOR S MANUAL 124 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 133: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...