Chapter 2: Basic Pump Operation
951128 Rev E (11/10)
13
timeframe. If the pump is not plugged into A/C power within the Low Battery Warning
timeframe, a Plug In Alarm may be displayed. The Plug In Alarm alerts you to the fact that
there is only 3 minutes of battery life remaining and you should immediately plug the device
into an A/C outlet. If the battery discharges completely, the infusion will stop and you will
need to plug the pump into an A/C outlet to restart the infusion and recharge the battery.
N
OTE
: The Low Battery Warning will be displayed when there is
no
less
than 30 minutes of battery life remaining. Additional battery
life is possible based on battery condition, delivery rate and other
external factors but it is recommended that the pump be plugged
into A/C power within the 30 minute timeframe to avoid shutdown.
PRIMING THE TUBING
W
ARNING
: Always read and follow the instructions which accompany
the source container and IV administration sets you are using.
Carefully follow the instructions for loading, removing and reloading
the set.
C
AUTION
: Administration sets should be changed at an interval not
to exceed 72 hours to maintain specified accuracy.
1. Close roller clamp on IV set.
2. Spike IV container and squeeze drip chamber until one-third or one-half full.
3. Prime the IV set by opening the roller clamp and letting fluid flow through the tubing with
the cassette inverted (the pointed end of the cassette should be up).
4. Ensure all air is expelled from the cassette and IV tubing, then close the roller clamp.
LOADING THE TUBING
C
AUTION
: Use only B. Braun Medical Inc. Horizon® Pump IV Sets.
1. Open the door on the front of the pump by pulling the Door Lever forward.
2. Position the cassette over the Cassette Alignment Pins, with the metal disc on the cassette
towards the pump.
3. Thread the tubing through the Air-In-Line Detector.
4. Insert and press the Free Flow Protection Clip into the Receptacle.
W
ARNING
: Unrestricted fluid flow may occur if the Free Flow
Protection Clip is not properly installed in the Receptacle.
Содержание OUTLOOK ES Series
Страница 2: ...951128 Rev E 11 10 ...
Страница 4: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 30: ...OUTLOOK ES OPERATOR S MANUAL 22 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 90: ...OUTLOOK ES OPERATOR S MANUAL 82 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 96: ...OUTLOOK ES OPERATOR S MANUAL 88 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 102: ...OUTLOOK ES OPERATOR S MANUAL 94 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 131: ...951128 Rev E 11 10 longer intervals of time Trumpet Curve 2nd Hour of Delivery Trumpet Curve Last Hour of Delivery ...
Страница 132: ...OUTLOOK ES OPERATOR S MANUAL 124 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 133: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...