O
UTLOOK
®
ES
O
PERATOR
’
S
M
ANUAL
16
951128 Rev E (11/10)
PUMP BEHAVIOR
The Outlook® ES behaves in any one of the following four ways.
Hold
In general, when
is pressed, the infusion stops. Pressing
will also silence an
alarm and stop the infusion. While in the menu system, pressing
will exit the user
back to the current mode and stop the infusion (if running).
N
OTE
: In certain cases during programming, pressing the HOLD
key will generate a pop up message that will instruct the user to
continue programming or return the pump to the last accepted
infusion parameters. This is a useful feature when infusion
parameters need to be corrected or reinput during the
programming sequence.
Run
When the Run Indicators are blinking, the infusion is running at the entered rate. Pressing
PRIMARY
or
PIGGYBACK
will begin an infusion as long as all required param-
eters are entered. If the
RATE
or
VTBD
(Volume To Be Delivered) fields have been left at
zero, the pump will alarm and prompt the user to enter a rate or volume to be delivered
value.
Alarm
The pump alarms when an error has occurred in the pumping process or when the Hold
Time has been exceeded. During an alarm, the pump stops the infusion and an audible
alarm sounds, different from the other audible tones of the pump.
KVO
Keep Vein Open (KVO) allows the pump to transition from its primary infusion to a low
delivery rate once the volume to be delivered value goes to zero. The pump will sound a
KVO alert tone which is different from the other alarm tones. Pressing
will silence
the KVO alert tone and stop the infusion. The KVO rates are as follows:
Infusion Rate < 3.0 mL/hr : KVO Rate = last infusion rate
Infusion Rate > 3.0 mL/hr : KVO Rate = 3.0 mL/hr
Содержание OUTLOOK ES Series
Страница 2: ...951128 Rev E 11 10 ...
Страница 4: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 30: ...OUTLOOK ES OPERATOR S MANUAL 22 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 90: ...OUTLOOK ES OPERATOR S MANUAL 82 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 96: ...OUTLOOK ES OPERATOR S MANUAL 88 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 102: ...OUTLOOK ES OPERATOR S MANUAL 94 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 131: ...951128 Rev E 11 10 longer intervals of time Trumpet Curve 2nd Hour of Delivery Trumpet Curve Last Hour of Delivery ...
Страница 132: ...OUTLOOK ES OPERATOR S MANUAL 124 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 133: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...