58
951128 Rev E (11/10)
If either the starting infusion rate or dose is 0.
0
, the
appropriate field will be highlighted when the scan
is complete. The following steps should be
completed to proceed with the infusion
programming sequence.
1. Enter a valid, non-zero value for rate or dose (as
applicable)
2. Use the
to highlight "Accept", then press
.
N
OTE
: The highlight cursor will return to the appropriate rate or
dose field if the 0.o value has not been changed to a valid, non-
zero value.
FOR LABELS WITH "0.
0
" FOR PATIENT WEIGHT
The DoseScan™ feature also has the ability to
accept a 0.
0
(zero) value for patient weight in those
cases where the patient weight has not been
specified. After completing the scanning sequence
described in the previous section, the following
validation screens will appear.
The following steps should be completed to proceed
with the infusion programming sequence.
1. Enter a valid, non-zero value for Body Weight.
2. Use the
to highlight "Accept", then press
.
N
OTE
: The highlight cursor will return to the weight field if the 0.o
value has not been changed to a valid, non-zero value.
TITRATION CHECKING
The DoseScan™ feature also has the ability to check scanned information with the previous
infusion parameters. This is especially useful in those circumstances where multiple bags of
the same medication are sent to the care unit for sequential delivery. Changes in rate or dose
could be an indication of a titration. Titration checking provides a means to confirm the
proper rate or dose for the scanned medication. The process is completed as follows:
Содержание OUTLOOK ES Series
Страница 2: ...951128 Rev E 11 10 ...
Страница 4: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 30: ...OUTLOOK ES OPERATOR S MANUAL 22 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 90: ...OUTLOOK ES OPERATOR S MANUAL 82 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 96: ...OUTLOOK ES OPERATOR S MANUAL 88 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 102: ...OUTLOOK ES OPERATOR S MANUAL 94 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 131: ...951128 Rev E 11 10 longer intervals of time Trumpet Curve 2nd Hour of Delivery Trumpet Curve Last Hour of Delivery ...
Страница 132: ...OUTLOOK ES OPERATOR S MANUAL 124 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 133: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...