O
UTLOOK
®
ES
O
PERATOR
’
S
M
ANUAL
54
951128 Rev E (11/10)
N
OTE
: Canceling the bolus will return the infusion to the continuous
infusion parameters and associated display.
EPIDURAL BOLUS CAUTION
The Micro Rate Limit feature is automatically invoked when as drug or solution is
programmed from the Epidural Care Area of the Drug Library. The Micro Rate Limit feature
limits the delivery rate of either the continuous infusion or bolus delivery to a maximum of
99.9 ml/hr. Your facility may have set a lower Micro Rate limit. It also automatically
increases the delivery pressure to 750 mmHg. If the programmed rate exceed the Micro Rate
limit a warning screen will be displayed similar to the screen shown below: