STANDARD Mode allows for infusion of fluids when rate and volume to be
delivered data is entered. For the Outlook® ES, this delivery is also known as a
Basic Infusion. At the end of the infusion, the pump automatically switches to
Keep Vein Open in order to maintain IV patency. In this mode, secondary fluid,
such as an antibiotic, may be administered using a PIGGYBACK method of
delivery. Piggyback infusions are allowed only in STANDARD Mode.
1. Press
PRIMARY CHANNEL INDICATOR
.
2. Use the
keys to enter the rate and volume to be
delivered.
3. Press
PRIMARY
to start the infusion. The green
Run Indicators will light up.
W
ARNING
: After pressing
PRIMARY
,
make sure drops are falling
in the drip chamber. If no drops are falling, make sure the roller
clamp is open. If the roller clamp is open and still no drops are
falling, remove and replace the set as it may be defective.
The Time Left, Total Infused, and Occlusion Limit are shown on the Information Screen.
Monitor the infusion according to hospital policy. After the infusion is complete, the pump
will alarm and go into KVO.
CHANGING THE RATE
With Pump Infusing
1. Press the
keys to change the rate.
2. Press
or
PRIMARY
to confirm new rate.
N
OTE
: If
or
PRIMARY
is not pressed within 4 seconds of
entering a new rate, a series of error beeps will occur and the rate
change will be cancelled. The pump will then display the message
“Rate change aborted,” and will continue to infuse at the old rate. An
intermittent alarm will continue to sound to notify the user that a rate
change has been attempted and cancelled. Pressing any key will
silence the alarm.
Infusion
KVO
RUN
125.
0
1000.
0
PRIMARY
PRIMARY
PIGGYBACK
PIGGYBACK
RATE ML/HR VOLUME TO BE DELIVERED ML
RUN
Содержание OUTLOOK ES Series
Страница 2: ...951128 Rev E 11 10 ...
Страница 4: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 30: ...OUTLOOK ES OPERATOR S MANUAL 22 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 90: ...OUTLOOK ES OPERATOR S MANUAL 82 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 96: ...OUTLOOK ES OPERATOR S MANUAL 88 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 102: ...OUTLOOK ES OPERATOR S MANUAL 94 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 131: ...951128 Rev E 11 10 longer intervals of time Trumpet Curve 2nd Hour of Delivery Trumpet Curve Last Hour of Delivery ...
Страница 132: ...OUTLOOK ES OPERATOR S MANUAL 124 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 133: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...