951128 Rev E (11/10)
N
OTE
: As battery life is dependent on delivery rate, it is
recommended that you plug the pump into an AC power source
when the battery life indicator displays a charge indication in the
yellow portion of the bar.
To Exit the
Battery Level
screen, press
.
CHANGE ALARM VOLUME
The volume of the pump alarm can be changed to
low, medium, or high.
1. Press .
2. Use the
to highlight “
Change Alarm Vol-
ume
,” then press
.
3. Use the up and down
to highlight
“
LOW
,” “
MEDIUM
,” or “
HIGH
.”
4. Press
to confirm the new volume setting.
SELECT MAX PRESSURE
The Outlook® ES can be used for epidural administration of anesthetic and analgesic drugs.
Use this feature when the maximum pressure of 750 mmHg is required for infusion.
1. Press
to stop the infusion.
2. Press .
3. Use the
to highlight “
Select MAX Pressure
,” then press
.
The Occlusion Limit is now set to 750 mmHg. To change the Occlusion Limit from 750
mmHg, turn the pump off and then on again, or See “
Occlusion Limit
” to change the value.
MICRO RATE/LIMIT
This option is useful in environments where low rates are critical. For patient safety, the
maximum rate limit can be set from 5.0 ml/hr to 99.9 ml/hr.
N
OTE
: If the pump is turned on with Micro
Rate/Limit activated, a teddy bear symbol
shows on the Information Display for 4
seconds.
To Purchase, Visit
Содержание OUTLOOK ES Series
Страница 2: ...951128 Rev E 11 10 ...
Страница 4: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...
Страница 30: ...OUTLOOK ES OPERATOR S MANUAL 22 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 90: ...OUTLOOK ES OPERATOR S MANUAL 82 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 96: ...OUTLOOK ES OPERATOR S MANUAL 88 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 102: ...OUTLOOK ES OPERATOR S MANUAL 94 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 131: ...951128 Rev E 11 10 longer intervals of time Trumpet Curve 2nd Hour of Delivery Trumpet Curve Last Hour of Delivery ...
Страница 132: ...OUTLOOK ES OPERATOR S MANUAL 124 951128 Rev E 11 10 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 133: ...951128 Rev E 11 10 To Purchase Visit Avobus com or call 1 800 674 3655 ...