![Axon Axopatch 200B Скачать руководство пользователя страница 88](http://html.mh-extra.com/html/axon/axopatch-200b/axopatch-200b_theory-of-operation_3040439088.webp)
72
•
GENERAL INFORMATION
Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.
4) Try grounding auxiliary equipment from a ground distribution bus. This bus
should be connected to the Axopatch 200B via the yellow banana (4 mm)
socket on the rear panel. This socket is connected to the signal ground of the
Axopatch 200B (
i.e.
, the outer conductors of all the BNC connectors). The
signal ground in the Axopatch 200B is isolated from the chassis and power
ground.
5) Experiment. While hum can be explained in theory (
e.g.
, direct pickup, earth
loops), in practice the ultimate theory is the end result. Following the rules
above is the best start. The final hum level can often be kept to less than
0.1 pA
p-p
. One technique that should
not
be used to reduce hum is the delicate
placement of cables so that a number of competing hum sources cancel out.
Such a procedure is too prone to accidental alteration.
Model Cell
The PATCH-1U model cell (Figure 13) can be used to assist with testing and setting up.
The pipette is modeled by a 10 M
Ω
resistor, the cell is modeled by 500 M
Ω
in parallel with
33 pF (the membrane time constant is 16.5 ms), and the patch is modeled by a 10 G
Ω
resistor. The pipette capacitance is about 4-6 pF. The charging time constant is
approximately 330 µs (10 M
Ω
x 33 pF).
The PATCH-1U model cell has been made without a switch to change the model between
BATH, PATCH and CELL positions. This is because even the best switches have an
enormous amount of leakage resistance and capacitance which increases the noise three to
five times beyond what you can achieve with a good seal. Instead of switches, three
separate plug positions have been provided and you can rotate the model cell into the
position required. With this technique the noise contribution of the model cell is still
somewhat more than can be achieved with a good seal, but the increase is less than 50%.
The PATCH-1U model cell can conveniently be used in conjunction with the tutorial at the
front of this manual.
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...