USE OF THE PATCH CLAMP – A TUTORIAL
•
15
Chapter 3
CONDITIONS:
PATCH CONFIG.
SEAL TEST ON 5kHz FILTER
Figure 2.
Electrode (pipette) capacitance compensation.
Whole-Cell Recording (Model Cell)
Set the front panel controls of the Axopatch 200B as follows:
PIPETTE OFFSET:
About 5.0
ZAP:
0.5 ms
SERIES RESISTANCE COMP. % PREDICTION:
0 %, OFF
SERIES RESISTANCE COMP. % CORRECTION:
0 %, OFF
SERIES RESISTANCE COMP. LAG:
1 µs
WHOLE CELL CAP.:
0 pF, OFF
SERIES RESISTANCE:
0 M
Ω
HOLDING COMMAND:
0 mV, toggle x1, OFF
SEAL TEST:
OFF
METER:
Set switch to V
TRACK
MODE:
TRACK
CONFIG.:
WHOLE CELL (
β
= 1)
OUTPUT GAIN:
α
= 1
LOWPASS BESSEL FILTER:
5 kHz
LEAK SUBTRACTION:
∞
M
Ω
, OFF
↑
↑
PULSE
PULSE
ON
OFF
1111 p
pp
pA
AA
A
1111 m
m
m
mssss