PRINCIPLES OF OPERATION
Axopatch 200B, Copyright 1997-1999, Axon Instruments, Inc.
Inward / Outward Current
In a cell generating an action potential, depolarization is caused by a flow of
positive sodium or calcium ions
into
the cell. That is,
depolarization
in this case is
caused by an
inward
current.
During intracellular current clamping, a depolarizing current is a
positive
current
out of the microelectrode tip into the interior of the cell. This current then passes
through the membrane
out
of the cell into the bathing solution. Thus, in
intracellular current clamping, a
depolarizing (positive)
current is an
outward
current.
An
inward
sodium current flows in some cells after a depolarizing voltage step.
When the cell is voltage clamped, the sodium current is canceled by an equal and
opposite current flowing into the headstage via the microelectrode. Thus it is a
negative
current. When two-electrode voltage clamping was first used in the early
1950's, the investigators chose to call the
negative
current that they measured a
depolarizing
current because it corresponded to the depolarizing sodium current.
This choice, while based on sound logic, was unfortunate because it means that
from the recording instrument's point of view, a negative current is
hyperpolarizing
in intracellular current-clamp experiments but
depolarizing
in voltage-clamp
experiments.
To prevent confusion, Axon Instruments has decided to always use current and
voltage conventions based on the instrument's perspective. That is, the current is
always unambiguously defined with respect to the direction of flow into or out of
the headstage. Some instrument designers have put switches into the instruments
to reverse the current and even the command voltage polarities so that the
researcher can switch the polarities depending on the type of experiment. This
approach has been rejected by Axon Instruments because of the real danger that if
the researcher forgets to move the switch to the preferred position, the data
recorded on the computer could be wrongly interpreted. Axon Instruments believes
that the data should be recorded unambiguously.
Содержание Axopatch 200B
Страница 2: ......
Страница 8: ...INFORMACION IMPORTANTE Axopatch 200B Copyright March 1997 1999 Axon Instruments Inc ...
Страница 26: ......
Страница 56: ......
Страница 86: ......
Страница 118: ......
Страница 134: ......
Страница 136: ......
Страница 146: ......
Страница 148: ......
Страница 150: ......
Страница 152: ......
Страница 153: ...Circuit Diagrams Request Form ...